General
The AlM-M200 monitors the insulation resistance of medical IT system, At the same time, the load current and temperature of the IT system transformer are monitored. In combination with ASG100 AIL100series insulation fault locators and the appropriate measuring current transformers, the AlM-M200 is designed to support insulation fault location.
Technical parameters
Model | AIM-M10 | AIM-M100 | AIM-M200 | |
Power supply | Voltage | AC220V±10% | AC220V±10% | DC24V±25% |
Frequency | 50/60Hz | 50/60Hz | - | |
Max power consumption | <5W | <8W | <3W | |
Insulation monitoring | Insulation resistance measuring range | 10-999kΩ | ||
Relative percentage accuracy | 0—±10% | |||
Alarm value range | 10—995kΩ | 50—999kΩ | 50—999kΩ | |
Response time | <2s | |||
Measuring voltage | <12V | |||
Measuring current | <42uA | <50uA | <50uA |
Current monitoring | Measuring range | 2.1-50A | ||
Alamm value range | 5-50A | |||
Measurement accuracy | ≤±5% | |||
Temperature monitoring | Thermal resistor | PT100 | ||
Measuring range | -50—+200℃ | |||
Alarm value range | 0—+200℃ | |||
Alarm output | Output type | 1 channel relay output | 2 channels relay output(programmable) | 2 channels relay output |
Contact rating | AC 250V/3A DC 30V/3A | AC 250V/3A DC 30V/3A | AC 250V/3A DC 30V/3A | |
Environment | Working temperature | -10—+55℃ | ||
Storage temperature | -20—+70℃ | |||
Relative humidity | 5%-95%,non-condensing | |||
Altitude | ≤2500m | |||
Communication | RS485 interface,Modbus-RTU protocol | RS485 interface,Modbus-RTU protocol | CAN interface, custom protocol, RS485 interface, Modbus-RTU protoco |