AIM-M300 series medical intelligent insulation monitor
AIM-M300 series medical intelligent monitoring device adopts advanced microcontroller technology, which has high integration, compact size, convenient installation and integrates intelligence, digitalization and networking in one. It is suitable for insulation monitoring of isolation power system in Class 2 medical locations 2 such as operating room and intensive care unit.
Technical Parameters
AUX Power | Voltage | DC 18-36V | Locate Signal | Locate voltage | <12V |
Power consumption | ≤6W | Locate current | ≤1mA | ||
Insulation Monitoring | Resistance range | 15-999kΩ | Alarm Output | Output mode | 2 output relay |
Response value | 50-999kΩ | Contact rating | AC 250V/3A; DC 30V/3A; | ||
Relative uncertainty | ±10% | Communication Protocol | Communication 1 | RS485, Modbus-RTU | |
Response time | ≤3s | Communication 2 | CAN, customize | ||
Permissible system leakage capacitance | ≤5μF | Environment | Operating temperature | -10~+55°C | |
Allowed external voltage Ufg | ≤DC 280V | Transport temperature | -25~+70°C | ||
Measuring voltage | ≤13V | Storage temperature | -25~+70°C | ||
Measuring current | ≤50μA | Relative humidity | 5-95%, No condensation | ||
lmpedance Zi | ≥240kΩ | Altitude | <2500m | ||
Current Monitoring | Measuring Value | 2.1-60A | Insulation method | Guide | |
Measuring Accuracy | ±5% | Display Method | LCD, LED | ||
Alarm Value | 5-80A | Fault Record | 20 | ||
Temperature Monitoring | Therrnal resistor | two Pt100 | IP degree | IP40 | |
Measuring range | 50~+200°C | Rated impulse voltage/ pollution degree | 4KV/III | ||
Alarm value range | 0~+200°C | EMC/EMR | IEC61326-2-4 |
Dimension and wiring
Terminals
Wiring Diagram
Front Panel
Manufacturer Company
Energy meters/Soultions/Current sensors, please contact us