Y09-310NW Laser Airborne Particle Counter High Resolution
28.3L/min touch screen Laser Airborne Particle Counter is a new style particle counter with big flow rate (28.3L/min), friendly interface and high-resolution color touch screen, easy to used. It's used for measuring the particle size and counting the number of the particles. The model y09-310NW can finish sampling a cubic meter quickly and exactly in 36 minutes, confirm the class of the clean room and print out the test result quickly, meet the ISO 14644 absolutely. The instrument monitor the quantity and the changes of the particles in 6 particle channels simultaneously, It's widely used in testing particles of clean room, researching the airborne particle, evaluating quality of the indoor air, it also used for testing the performance of the filter and judging purify class
The specification of portable particle counter
Outside Dim | 245*268*160 (mm3); (Wide*Deep*High) |
Display | 7 inch touch screen |
Weight | 3.5Kg |
Max. Consumption | 80W |
Power Source | DC16.8V built-in Lithium Battery; AC 220V ± 10% |
Size Channels | 0.3, 0.5, 1, 3, 5, 10 μm |
Calculation | 95% UCL calculation, can display and print particle concentration (particles / cubic meter, particles /cubic feet) |
Flow Rate | 28.3L/min (Imported Pump) |
Sample Time | 6s---999999s (Selectable) |
Cycles | 0-99 |
Hold Time | 0-9999s |
Delay Time | 5-999s |
Zero Court | <=10min |
Location | 1-99 |
Alarm | ISO Class |
Printer | Built-in Thermal Printer |
Data Storage | 20000Samples /300 Samples of UCL (With PC Software) |
Light Source | Laser Diode (30,000 hours Mean Time Between Failures) |
Counting efficiency | 50% @ 0.3μm and 100% for particles >0.45μm (ISO21501) |
Maximum Rated Particle Concentration | 35000 particles/L |
Operating Time | More than 5.5 hours |
Environmental Temperature | Temperature :10 -35℃; Humidity:20%-75%RH; Atmosphere Pressure:86Kpa-106Kpa |
Features of air particle counter
1. Sample volume 28.3 LPM (1.0CFM).
2. Simultaneously display of up to six particle sizes
3. 20000-record data storage memory
4. Meet the requirements of ISO14644-1:2015
5.Built-in Thermal Printer
6.Low power display
SUJING