Leakage Tracking Tester LT-3 0~600V Adjustable Precision 1.5%
Standard of Leakage Tracking Tester:
IEC60112-2009, UL746A, IEC60335-1, IEC60598-1 and ASTM D3638-92.
Application of Leakage Tracking Tester:
It is used to evaluate the tracking resistance performance of solid insulating material surface under the combined influence of electric field and humid or contaminate medium. In a word, this device is used to measure the compare tracking index (CT1) and electrical resistance index (PT1).
Test Sample of Leakage Tracking Tester:
Lighting equipment, low-voltage electrical appliances, household electrical appliances, machine tools, electrical appliances, electric motors, electric tools, electronic instruments, electrical instruments, information technology equipment, insulation materials, engineering plastics, electrical connectors, accessories industry.
Feature of Leakage Tracking Tester:
The working principle of the leakage tracking test (tracking index testing) is that the conducting liquid (0.1%NH 4 CL) of required volume in the required height (35mm) and required time (30s) drops with the voltage between the platinum electrodes (2mm× 5mm) on the surface of solid insulating material. Thus, the users evaluate the tracking resistance performance of solid insulating material surface under the combined influence of electric field and humid or contaminate medium. In a word, this device is used to measure the compare tracking index (CT1) and electrical resistance index (PT1).
Parameter of Leakage Tracking Tester:
Model | LT-3 (0.5m³) (Touch screen) |
Working voltage | 220V/50Hz,1KVA |
Control operation mode | PLC control, 7 inches color touch screen operation, with infrared remote control |
Testing voltage | 0~600V adjustable,precision 1.5% |
Timing device | 9999X0.1S |
Electrode | Material:Platinum electrode and brass connecting rod |
Size:(5±0.1)×(2±0.1)×(≥12)mm,30°slant,Tip rounding:R0.1mm | |
Relative position of electrode | Included angle:60°±5°, distance is 4±0.1mm |
Electrode pressure | 1.00N±0.05N,Touch screen display |
Dripping liquid | The interval time of dropping liquid:30±5S,digital display,can be preset |
Height:35±5mm | |
The number of drips:0-9999 times,can be preset, the volume size of the dripping liquid is controlled by the imported micro pump within 50 ~ 45 drips /cm³ | |
Test liquid resistance | A liquid 0.1%NH4Cl,3.95±0.05Ωm, B liquid 1.7±0.05Ωm |
Time-delay circuit | 2±0.1S (in 0.5A or larger current) |
Short-circuit pressure drop | 1±0.1A 1%,pressure drop 8% MAX |
Wind speed | 0.2m/s |
Environmental requirement | 0~40℃,relative humidity≤80%,in the place of no obvious vibration and corrosive gas |
Dimensions and weight | WDH=1100×550×1200 mm,about 130kg |