ECG007oem module spo2 oem module ECG module
Technical Specifications:
1. Principle of Operation: Non-dispersive infrared (NDIR)
2. Initialization Time: Capnogram displayed in less than 20 seconds Reaching full specifications
within 2 minutes.
3. Co2 Measure Range: 0 - 150 mmHg; 0 - 19.7%; 0 - 20kPa
Co2 Measure Resolution: 0.1mmHg
Co2 Accuracy: 0-10mmHg ± 2mmHg; 41-70mmHg ± 5% of reading;
71-100mmHg ± 8% of reading; 101-150mmHg ± 10% of reading
4. Calibration: No routine user calibration required, The mainstream sensor an airway adapter
5. Zero is required when changing to a different airway adapter
6. Compensation: Barometer Pressure 450 - 850 mmHg.
7. Operator selectable O2, N2O, He compensation.
8. Data Interface: Rs232, 19200 baud rate, standard N-8-1.
9. Regulatory: Designed to comply YY0601 and ISO21647
The YS3000 is a stand-alone external pulse oximetry module that can be externally connected to a multiparameter monitor,or a hand-held oximeter and other devices that require measurement of oxygen saturation. Connecting to your device via a simple TLL interface, the module can measure blood oxygen saturation accurately with a SpO2 sensor.
► Small volume,flexible installment mode,high reliable and accurate test. |
► Excellent reliability and clinical monitoring fuction. |
► It has implemented the monitor of 7 channel ECG, 1channel RESP and2channels temp on the same time. |
► In accordance with function safety design of CE requirements. |
► In accordance with all the ecg/temp monitoring work |
► Suitable for adult,pediatric and neonate. |
► Single power 5V work,low power consumption design. |
Feature
1. Small volume,flexible installment mode,high reliable and accurate test.
2. Excellent reliability and clinical monitoring fuction.
3. It has implemented the monitor of 7 channel ECG, 1channel RESP and2channels temp on the same time.
4. In accordance with function safety design of CE requirements.
5. In accordance with all the ecg/temp monitoring work,
6. Suitable for adult,pediatric and neonate.
7. Single power 5V work,low power consumption design.
Specification
ECG:
Full lead: I, II, III, aVL, aVR, aVF and V
Gain 5mm/mV, 10mm/mV, 20mm/mV and 40mm/mV
Scanning speed: Changeable with three grades of 12.5mm/s, 25mm/s and 50mm/s
Band width 0.05Hz-100Hz(3dB)
Heart rate:
Scope: 1 times/minute to 300 times/minutes
Accuracy: ±1 time/minute
Variation of heart beat LORENZE drawing:Analysis on heart beat data for the
current 5 minutes with histogram during RR period.
Resp
Chest impedance method or concentration of carbon dioxide at end of respiration
Scope of chest impedance method 0 time/minute – 100 minutes/minute
Precision of chest impedance method 1 time/minute
Detection scope of carbon oxygen at end of respiration 3mmHg-75mmHg; 0.4%vol-9.9%vol
Detection scope of respiration rate: 2 times/minute ->60 times/minute
(including baby)
Detection precision of respiration rate: ±1 time/minute
Temp
Single/double path
Type: Vascular/surface temperature
Scope:20℃~50℃
Accuracy±0.2℃