BLX - 6 Low Dose Dental X - ray Unit Portable Dental Panoramic X - ray Machine
Specifications:
1) Power(V): AC220 50Hz |
2) Tube Voltage: 60KV |
3) Tube Current: 0.1MA |
4) Frequency: 30KHz |
5) Time of Exposure: 1-9s |
6) Radiation Leakage: <20μgy/h |
7) Rated Power: 60W |
8) Weight: 2KG |
9) Environmental Temperature: 5-40°C |
10) Relative Humidity: <80% |
Packing List:
1. Main Unit
2. Ball Head
3. 2 cables
4. Other accessories for test. (Kodak Films,etc.)
Reference of the Angles:
Tooth | Upper | Lower |
1-2 | 42 | -15 |
3 | 45 | -18 |
4-5 | 30 | -10 |
6-8 | 28 | -5 |
Reference of the time of exposure:
Tooth | Adult | Child |
Central & Lateral | 2.5-3.25 | 2-2.7 |
Cuspid & 1st Bicuspid | 3.2-4 | 2.7-3.5 |
Upper 5,6 | 3.5-4 | 3-3.5 |
Upper 7,8 | 7.5-8 | 7-7.5 |
Lower 5,6 | 4-4.5 | 3.5-4 |
Lower 7,8 | 4-4.5 | 3.5-4 |
Trouble Shooting Instructions:
Failure | Reason | Solution |
Machine does not display after being connected to power | Loose connection of the plug | Check the plug again and the connection of the joints |
Broken fuss | Change the fuss | |
The pilot lamp lights up but the digital tube Of the main unit does net show the fore setting time | Abnormal Power supply | Plug to the right Power supply |
Circuit damaged | Send the product back to the factory or distributor to repair | |
No image is taken while the counting down works | Connection problem with the Main unit and the Ball head | Send the product back to the factory or distributor to repair |
The remote control does not work Within the normal distance | Battery low | Change for new batteries |
Damage in the remote control | Change for a new remote control or sent it for a repair | |
The time setting does not work | Error in the control program | Re-start on the machine |
Problem with the SET button | Send it for a repair | |
No counting down after press Start | The pilot lamp light is on for the SET | Shut down the SET mode |
Blurred film | The film,tooth,and ball head are not stable while taking a picture | Operate again rightly |
The film is too far from the tooth that is taken a picture | Operate again rightly | |
Wrong angle of taking a picture | Operate again rightly | |
The image is too light | The time of exposure is too short | Reset the exposure time and make it longer |
The time of developing the film is too short | Lengthen the developing time | |
The image is too dark | The time of exposure is too long | Reset theexposure time and make it shorter |
The time of developing the film is too long | Shorten the developing time | |
Other possible reasons for a Blurred film or a blank film | The film is of poor quality | Change for a better film |
The imaging liquid is of poor quality | Change for a better liquid | |
The temperature is too low while developing the film | Develop the film In a higher temperature |
Notice:
1. For your safety, do not open the unit. Only professional personnel can inspect and repair the vacuum components and high-voltage parts of the X-ray machine.
2. Tubehead equipped with perfect radiation-shield protection except the radial mouth. The test shows the radial index is zero. Safety is guaranteed.