Specifications
Brand Name :
Indair
Model Number :
CLJ-BII(D)
Certification :
ISO
Place of Origin :
China
MOQ :
1
Price :
negotiable
Supply Ability :
200 UNIT PER DAY
Packaging Details :
carton with plastic bag
Flow rate :
0.1 cfm (2.83 lpm)
Delay Time :
0-99sec
Self Clean Time :
≤10min
Location Labels :
0-999
Battery time :
8 hours
Max.Consumption :
80W
Description

Clean Room Handheld Laser Airborne Particle counter for Lab instrument

Handheld Particle Counter is used to detect the size and quantity of particles in the air per unit volume of the clean environments. It can quickly check or validate ISO Class 5-9 clean rooms and controlled environments.

With LCD display, the instrument uses laser diode as the light source. It has many features, such as slim design, light weight, high precision, simple operation, micro computer control, and can store and print sampling results. It is widely used in the electronics, optics, chemistry, food and beverage, pharmacy and sanitation, biology, precision machinery, aerospace, and other fields.

Outside Dim 300*360*126 mm3(Wide*Deep*High)
Display LCD
Weight 0.6kg
Battery (Lithium Ion Battery),7.2V,2200mAh
Power Source Li-ion charger ,8.4V DC,1A:9-250V AC, 50-60Hz
Size Channels 0.3, 0.5, 1, 3, 5, 10 μm
Flow Rate 2.83L/min
Sample Time 1s-99h
Cycles 0-99
Hold Time 0-999sec
Delay Time 0-99sec
Light source Laser Diode
Zero Court <=10min
Alarm Low Battery ;100, 1000, 10000 Class /ISO
Coincidence Loss 5%,2000000ps/cf
Operating Time more than 5.5 hours
output RS-232 to pc or printer via optional cable
Location Labels 0-999
Data Storage 1000 Samples (With PC Software)
Environmental 10degrees celsius-40degrees celsius ; Humidity: 20%-90% RH ;Atmosphere Pressure 86Kpa-106Kpa
Standard feathure suitcase ;charger;sampling head;line of communication;communication software ;Automatic Purifier
Optional component External printer

Parameter:

l 8 particle-size channels: 0.3, 0.5, 0.7, 1.0, 2.0, 3.0, 5.0, 10.0(μm)

l Sample Period : 1 Sec~3599 Sec(5L-3000L)

l Self-clean Time : ≤10min

l Operating Environment : Temperature 10~35°C,Relative Humidity ≤75%

l Output : Built-in Printer

l Light source : Laser diode (30,000 hours mean time between failure)

l Max permissible sampling concentration :35.000PCS/L

l Dimension (W×L×H):215×500×240(mm)

l Power Rating : 250W

l Weight : 8KG

l Power Supply : 220V±10%,50Hz±2Hz

l Data conversion (ft3 and m3)

l Accessory: Case, Operation manual, Factory inspection report, Filter, Power line, Tripod ,Printing paper, Fuse


Clean Room Electronic Particle Counter High Sensitivity Laser Diode As Light Source

Send your message to this supplier
Send Now

Clean Room Electronic Particle Counter High Sensitivity Laser Diode As Light Source

Ask Latest Price
Brand Name :
Indair
Model Number :
CLJ-BII(D)
Certification :
ISO
Place of Origin :
China
MOQ :
1
Price :
negotiable
Contact Supplier
Clean Room Electronic Particle Counter High Sensitivity Laser Diode As Light Source

Suzhou Indair indoor air technology co.,ltd

Active Member
7 Years
jiangsu, suzhou
Since 2004
Business Type :
Manufacturer, Distributor/Wholesaler, Exporter
Total Annual :
14700000-20000000
Employee Number :
100~500
Certification Level :
Active Member
Contact Supplier
Submit Requirement