Rubber Ozone Aging Resistance Test Chamber, Ozone Environmental Test Chamber
Application:
Rubber Ozone Aging Resistance Test Chamber, Ozone Environmental Test Chamber is used to test the ozone environment, the characteristics of crack-resistant rubber, ozone resistance testing machine, can control and adjust the concentration of ozone environment, to determine the anti-ozone properties of the specimen. The equipment can comply with the standard: JISK6259 ASTM1149 ISO1431 GB/TTB62 GB/T13642
Ozone Environmental Test Chamber Technical Parameters:
Model | OA-80 | OA-408 | OA-800 |
Internal Dimensions WxHxD (mm) | 400x500x400 | 600x850x800 | 1000x1000x800 |
External Dimensions WxHxD (mm) | 950x1650x950 | 1200x1950x1350 | 1600x2000x1450 |
Temperature Range | RT+10℃~+65℃ | ||
Control Accuracy | ±0.5℃ | ||
Distribution Uniformity | ±2.0℃ | ||
Analytic Accuracy | 0.1℃ | ||
Ozone Concentration (pphm) | 1~1000, adjustable | ||
Deviation of Ozone Concentration | ≤10% | ||
Ozone Generator | Silent discharge tube type | ||
Internal and External Material | Material of the inner box is SUS 304# stainless steel, of the outer box is stainless steel or see cold-rolled steel with paint coated. | ||
Insulation Material | Polyurethane rigid foam / superfine glass fiber cotton | ||
Security | Leakage, short circuit, over-current protection | ||
Power | AC 220V |
Rubber Ozone Aging Resistance Test Chamber Features:
1. Stainless steel inside and outside
2. Outside made iron with powder coating
3. Durable ozone generator
4. Insulation material is High-density glass fiber cotton
5. Bottom is universal round wheel and fixed foot
6. Window materials is multilayer steel-glass, the inside is chip conductive
Safety protection device:
1. the refrigeration system: The compressor overheating/ Compressor flow/Compressor overpressure/ Cooling fan overheating/ The compressor oil pressure protection
2. test chamber: Adjustable overtemperature/Air conditioning channel overtemperature limit/ Fan motor overheating.
3. the other: The total power phase sequence and the lack of protection /Electric leakage protection/Dry heating protection/ Overload and short circuit protection/ Overtemperature protection power.
Picture display function
1. In scene dialogue, no key presses, direct touch screen option.
2. Set the temperature and humidity (SV) and the actual value (PV) display directly.
3. Ozone concentration. Set (SV) and the actual value (PV) display directly.
4. Running total time accumilate function.
5. Temperature and humidity settings to program graphical curve shows a real-time curve display program execution function .
6. English and Chinese language available.
7. Troubleshooting tips showed.
8. The screen can be adjusted for backlight.
9. Screen protection for timing, closed set the TIMER or manually.
Certificate of these the machine:
China official 3rd party calibration report
CE certiification