Plasma Surgery System for Arthroscopy Surgery
Overview
The Plasma Surgery System for Arthroscopy Surgery is a cutting-edge medical device designed to enhance the precision and efficiency of arthroscopic procedures. Utilizing advanced plasma technology, it delivers a controlled energy source that allows for precise tissue ablation and coagulation. This system minimizes thermal damage to surrounding tissues, ensuring a safer and more effective treatment. Its ergonomic design and user-friendly interface make it easy for surgeons to navigate tight joint spaces with ease. Compatible with a wide range of arthroscopic instruments, this system reduces procedural time and promotes faster patient recovery. It is an indispensable tool for modern orthopedic surgeons, offering a reliable and innovative solution for joint surgeries.
Technical Parameter
Input Volts | 198-240V 50Hz 100-120V 60Hz Export device could convert automatically per customer requirements |
Output Power | Ablation or resection: ≤351W Coagulation or hemostasis: ≤111W |
Working Frequency | 100 KHz ± 10% |
Dimension | 405*440*130mm (L*W*H) |
Working Mode | Ablation and resection: level 1-10 Coagulation: level 1-10 |
Weight | 7.0kg |
The Revolution of Plasma Surgery
The low-temperature surgical system generates high-frequencyenergy in the working surface,therebyintegrated therapeutic func-tions of ablation,cutting,coagulation hemostasis,infusion,andsuction,which have changed traditional surgical methods,makingthe surgery can be performed without the need for additionalspecial instruments during the surgical process,in this way to com-plete the surgery quickly and efficiently.
Combining the characteristics of human physiological struc-ture for treatment;
Submucosal ablation effectively protects the mucosa,andmaintains the original physiological barrier,and avoids dam-aging the original physiological immunity;Real time ablationimmediate retraction of hypertrophic tissue during surgery.
Technical Principle
The plasma technology of Better Health is the plasmaenergy generated by stimulating NaCl in physiological salinewith a 100KHz square wave.The high-speed charged particlesin the plasma can break the molecular bonds of the humanbody,causing proteins and other tissues to decompose andvaporize into low molecular weight gases such as H2,02,CO2,N2,and methane.
Under a stable electric field of 100KHz square wave,parti-cles will achieve a longer acceleration time,ultimately forminghigh-speed charged particles with kinetic energy,which canbreak molecular bonds and cause minimal damage tosurrounding tissues.The treatment temperature is low,andthe processes of cutting,ablation,and hemostasis are com-pleted within 40-70℃.