Veterinary Ultrasound/veterinary Portable Ultrasound Scanner/cattle Pregnancy
1. Technical Specification
The main functions of the main probe
1. The frequency of Electronic Convex Array probe is 3.5MHz(R=60)
2. Depth display: ≥190mm
3. Resolution (vertical direction): ≤1mm (depth ≤130mm)
≤2mm(130mm<depth ≤160mm)
4. Resolution (cross direction) ≤2mm (depth ≤130mm)
≤3mm(130mm<depth ≤160mm)
5. Precision of geometric position :cross direction≤10%; vertical direction≤ 5%
6. Blind Zone(mm) ≤4mm
7. The conversional frequency points of the main probe (3.5MHz Convex Array probe) are: 2.0MHz﹑3.5 MHz﹑4.5 MHz﹑5.0 MHz﹑5.5 MHz﹑6.5 MHz﹑7.5 MHz , the detailed requirements of the conversion frequency points are as follows:
The frequency of probe( MHz) | f≤2.5 | 2.5<f≤4.0 | 4.0<f≤5.0 | 5.0<f≤10 | |
Type of probe | liner array , R≥60mm convex array | liner array , R≥60mm convex array
| liner array , R≥60mm convex array
| Superficial R≥60mm convex array
| (intra-cavity probe ) R<60mm convex Array) |
Resolution( lateral ) mm | ≤3 (depth)≤130) ≤4(130<depth≤160) | ≤2 (depth≤130) ≤3 (130<depth≤160) | ≤2 (depth≤60) | ≤1 (depth≤60) | ≤1 (depth≤40) |
Resolution (vertical direction)mm | ≤1 (depth≤130) ≤2 (130<depth≤170) | ≤1 (depth≤130) ≤2 (130<depth≤170) | ≤1 (depth≤80) | ≤1 (depth≤80) | ≤1 (depth≤40) |
Blind zone,mm | ≤4 | ≤4 | ≤3 | ≤3 | ≤5 |
Utmost depth display,mm | ≥190 | ≥190 | ≥120 | ≥80 | ≥60 |
Precision of geometric position,% | Lateral≤10 Vertical direction≤5 | Lateral≤10 Vertical direction≤5 | Lateral≤10 Vertical direction≤5 | Lateral≤5 Vertical direction ≤5 | Lateral≤10 Vertical direction≤5 |
2. The optional probes
1. 6.5MHz(R=13) Intra-cavity probe : 5.5 MHz﹑6.5 MHz and 7.5 MHz.
2. 7.5MHz Linear surface probe : 7.5 MHz﹑8.5 MH and 10.0 MHz.
3. Power
When AC 220V±22V or 110V±11V is changed , the equipment can also be operated as usual.
4. Working time
The continental working time for the Ultrasonic Diagnosis Equipment is more than 8 hours.
5. Selective configurations
The medical image work station can be connected with the ultrasonic if the function index of the former satisfies the requirements of the standard.
6. Safety
There are 60 items of Security Detection Project adapts
to the portable Ultrasound in the appendix A.
7. Environmental adaptation
It shall satisfy all the requirements of Climate environmental testing group Ⅱ and Mechanical environmental group Ⅱwhich are specified in the GB/T14710.