Dental Equipment 20: 1 Reduction Contra Angle Dental Implant
Surgical Low Speed Handpiece
Dental surgical implant handpieces technical data
Product name | Dental surgical implant handpiece |
Gear ratio | 20:1 |
Maximum speed | 0-2000rpm |
Max torque | 70 N·cm |
Working air pressure | 0.3~0.35Mpa |
Noise | ≤70db |
Bur applicable | Φ2.33~2.35mm |
Water irrigation | Internal and external irrigation |
Dental surgical implant handpieces features:
Cartridge , cap, transimisson shaft. fully compatible with W&H implant handpiece
|
ISO standard e-type connection Match up with any kinds of dental implant mot |
external and internal irrigation Max. speed 2000rpm
|
After each patient use , maintenance should be done as follows:
1. First remove the diamond burs , and clean up all the teeth on dental handpieces precipitate
2. Use a metal needle to clean the spray holes and fumaroles
Note: The spray holes and diamond burs into 40 -degree angle , fumaroles and diamond burs to a 30 degree angle
3. Place the clean lubricating oil is sprayed into the axial hole 1-2 seconds ( diamond burs is plugged holes )
4. In the bottom of the handpiece were ( with / without quick connector ) two kinds of styling , make the right amount of lubricant is injected into the bottom
5. The handpiece with quick connector combination ( first determine the water switch is turned off )
6. Replace the diamond burs , surrounded by white silk cloth handpiece head ( make sure the needle does not touch the diamond burs and white gauze , so as not to tie )
7. pedal depressed for about 20 seconds , look at whether there is in white gauze on any stain. If there are stains on the white gauze , again, add oil , starting from item 4 continued routine maintenance , know that white gauze until no stain
8. Finally, no more than 135 degrees Celsius temperature and high pressure sterilization