High Speed Dental Handpieces E-Generator Dental LED
High Speed Handpiece 2/4 Hole
High Speed Handpiece Led Features:
The DORIT LED high speed handpiece & swivel with inbuilt generator creates LED light in the mouth.
Simple FG bur changing
Precision bur cooling – triple coolant spray produces a mist around the bur.
Patented hygienic head – Prevents aerosol particles from being sucked into the turbine.
1 year warranty.
High Speed Handpiece Led Tech data:
Product name | High Speed Handpiece Led |
Drive air pressure | 0.25-0.3Mpa |
Water pressure | 0.2Mpa |
Water pressure | 198Kpa(2kg) |
Light brightness | 25500lux |
Power | 22W |
Rotation speed | 350,000~420,000rpm |
Head height | 13.0mm |
Head diameter | Φ11mm |
Chuck type | Push button |
Clamping force | 22N~34N |
Torque force | 0.1N·cm(small head); ≥0.1N·cm(large head) |
Bur Applicable | Φ1.59mm-Φ1.60mm*21mm-23mm(diameter*length) |
Noise | ≤60DB |
High Speed Dental Handpiece Led Detailed photos:
After each patient use , maintenance should be done as follows:
1. First remove the diamond burs , and clean up all the teeth on dental handpieces precipitate
2. Use a metal needle to clean the spray holes and fumaroles
Note: The spray holes and diamond burs into 40 -degree angle , fumaroles and diamond burs to a 30 degree angle
3. Place the clean lubricating oil is sprayed into the axial hole 1-2 seconds ( diamond burs is plugged holes )
4. In the bottom of the handpiece were ( with / without quick connector ) two kinds of styling , make the right amount of lubricant is injected into the bottom
5. The handpiece with quick connector combination ( first determine the water switch is turned off )
6. Replace the diamond burs , surrounded by white silk cloth handpiece head ( make sure the needle does not touch the diamond burs and white gauze , so as not to tie )
7. pedal depressed for about 20 seconds , look at whether there is in white gauze on any stain. If there are stains on the white gauze , again, add oil , starting from item 4 continued routine maintenance , know that white gauze until no stain
8. Finally, no more than 135 degrees Celsius temperature and high pressure sterilization