feeding pump
Technology Specification
Model |
CLS-300 Enteral feeding Pump |
Flow rate range |
1~400 ml/h |
Volume to be infused (VTBI) |
0 ~ 9999 ml |
Volume Infused (∑) |
0 ~36000 ml |
Infusion accuracy |
±10% |
Applicable feeding bag |
Support wide variety brands of feeding bag |
Bolus rate |
400 ml/h |
Mode of operation |
continuous |
Air bubble detection |
Ultrasound sensor detection with four levels of air in line alarm adjustable |
Occlusion pressure |
3 adjustable occlusion pressure settings: low, middle and high |
Alarms |
Visual and audible alarms: Door open, Air-In-Line, Occlusion, Infusion completion, infusion near over, Start reminder function, Low Battery, Battery depleted, malfunction etc. |
Computer Interface |
RS232 (optional) |
History records |
2000 history records |
Power supply |
AC:100~240V,50/60Hz DC:12V ±1V |
Battery |
Rechargeable lithium polymer battery,7.4V,2100mAh |
Battery recharge time |
10hs recharging when machine is turned on, 3hs recharging when machine is off. |
Power consumption |
25VA |
Classification |
Class I, type CF |
Water Proof Level |
IP×3 |
Mode of operation |
continuous |
Dimensions |
145×100×120 mm(length×width×height) |
weight |
≤1.4kg |
Operating conditions |
Temperature:5°C~40°C,relative humidity:10~95%, |
Transport conditions |
Temperature-20°C~+60°C,relative humidity:10~95%, |
Storage conditions |
Temperature-20°C~+45°C,relative humidity:10~95%, |