Specifications
Model Number :
TD1600
Name :
Fluke Multifunction Calibrator
Features :
wide range, multi value and high precision
Accuracy :
0.02% ; 0.05%
Application :
Clamp ammeters
Origin :
China
Model :
TD1600
Place of Origin :
China
Brand Name :
TUNKIA
MOQ :
negotiation
Price :
negotiation
Packaging Details :
Standard
Payment Terms :
negotiation
Description
TD1600 Precision AC Calibrator

Introduction

  • TD1600, a single-phase AC calibrator with wide range, high precision and multifunction.

  • The apparatus can output high accuracy AC voltage, current and power, and the optional accuracy are 0.05% and 0.02%. Applied to calibrate various AC electrical measuring instruments.

Features

  • AC Electrical Parameter Output: Voltage is up to 1100 V, current is up to 110 A.

  • Frequency and Phase Adjustment: Frequency: 45 Hz ~ 70 Hz.

  • AC Power Output: Measure the active power, reactive power, apparent power and power factor.

  • Value Output Mode: Standard source and output source modes, easy to calibrate both digital meters and analogue meters.

  • Value Adjustment Method: Fixed-point, rotary encoder, step and potentiometer (option).

  • Output Switch: The output can be turned on/off arbitrarily by one-touch operation.

  • Automatical Error Calculation: Set the measured values of the checked meter, and adjust the value output to automatically calculate the errors of the checked meter.

  • Excellent Human-computer Interaction: Large LCD touch screen display, touch and button operation modes, easy for users to measure manually.

  • Harmonic (Option): Load 2nd to 31st harmonics with variable amplitude and phase for impact test.

  • Proprietary Software (Option): Support the calibration of the checked meters by semi-automatic or full-automatic, data management and report export.

Specification

AC Voltage Output

Range Stability (%/min)

Accuracy (k = 2)

(ppm*RD + ppm*RG)

Max. Load Current

(mA)

0.05% 0.02% 0.05% 0.02%
1 V 0.01 0.005 300 + 200 120 + 80 500
3 V 0.01 0.005 300 + 200 120 + 80 500
10 V 0.01 0.005 300 + 200 120 + 80 500
30 V 0.01 0.005 300 + 200 120 + 80 500
60 V 0.01 0.005 300 + 200 120 + 80 800
100 V 0.01 0.005 300 + 200 120 + 80 500
150 V 0.01 0.005 300 + 200 120 + 80 400
200 V 0.01 0.005 300 + 200 120 + 80 250
300 V 0.01 0.005 300 + 200 120 + 80 150
400 V 0.01 0.005 300 + 200 120 + 80 100
600 V 0.01 0.005 300 + 200 120 + 80 80
750 V 0.01 0.005 300 + 200 120 + 80 60
1000 V 0.01 0.005 300 + 200 120 + 80 50

AC Current Output

Range Stability (%/min)

Accuracy (k = 2)

(ppm*RD + ppm*RG)

Max. Inductive

Load (V)

0.05% 0.02% 0.05% 0.02%
1 mA 1 μA 0.5 μA 300 + 2 μA 120 + 1 μA 120
2 mA 1 μA 0.5 μA 300 + 2 μA 120 + 1 μA 120
5 mA 1 μA 0.5 μA 300 + 2 μA 120 + 1 μA 120
10 mA 0.01 0.005 300 + 200 120 + 80 120
20 mA 0.01 0.005 300 + 200 120 + 80 120
50 mA 0.01 0.005 300 + 200 120 + 80 120
100 mA 0.01 0.005 300 + 200 120 + 80 120
200 mA 0.01 0.005 300 + 200 120 + 80 120
500 mA 0.01 0.005 300 + 200 120 + 80 100
1 A 0.01 0.005 300 + 200 120 + 80 50
2 A 0.01 0.005 300 + 200 120 + 80 25
5 A 0.01 0.005 300 + 200 120 + 80 10
10 A 0.01 0.005 300 + 200 120 + 80 5
20 A 0.01 0.005 300 + 200 120 + 80 2.5
30 A 0.01 0.005 300 + 200 120 + 80 2.5
50 A 0.01 0.005 300 + 200 120 + 80 1.5
100 A 0.01 0.005 300 + 200 120 + 80 0.8

Output Characteristics of Voltage and Current

  • Output Range of Voltage: 100 mV ~ 1100 V

  • Output Range of Current: 0.1 mA/ 20 mA ~ 110 A

  • Range Switching: Automatic or Manual.

  • Resolution: 0.001 %*RG; Display: 6 digits; Distortion: <0.2 %

  • 1 mA,2 mA,5 mA are small range option.

Frequency, Phase and Harmonic (Option)

Type Range Resolution Accuracy
Frequency 45.000 Hz ~ 70.000 Hz 0.001 Hz 0.005 Hz
Phase 0.0000 ° ~ 359.9999 ° 0.005 ° 0.05 °
Type Order Amplitude Variable Phase
Harmonic 2nd to 31st 0 ~ 25 % 0 ~ 359.9 °

Power Output

Type Stability (%/min) Accuracy (k = 2) %*FS
0.05% 0.02% 0.05% 0.02%
Active Power 0.01 0.005 0.05 0.025
Reactive Power 0.02 0.01 0.1 0.05
Apparent Power 0.02 0.01 0.1 0.05
Power Factor 0.02 0.01 0.1 0.05
  • Output Range of AC Power: Combination of AC voltage range and AC current range.

  • Range of Power Factor: -1.0000…0.0000…+1.0000

  • Active Power (|cosφ| ≥0.5), Reactive Power (|sinφ| ≥0.5)

Send your message to this supplier
Send Now

TD1600 Fluke Multifunction Calibrator High Precision Single Phase AC Calibrator

Ask Latest Price
Model Number :
TD1600
Name :
Fluke Multifunction Calibrator
Features :
wide range, multi value and high precision
Accuracy :
0.02% ; 0.05%
Application :
Clamp ammeters
Origin :
China
Contact Supplier
TD1600 Fluke Multifunction Calibrator High Precision Single Phase AC Calibrator

Tunkia Co., Ltd.

Verified Supplier
2 Years
Since 1998
Business Type :
Manufacturer, Seller, Other
Main Products :
Employee Number :
300~500
Certification Level :
Verified Supplier
Contact Supplier
Submit Requirement