ENT treatment unit is a integrated equipment which made according to the needs of patients and hospitals. It has curved and straight spray which can bring the liquid medicine atomization. The suction device used for sucking the phlegm and blood.
Subsidiary treatment table,Size | 1655mm*730mm*885mm 1pc |
Compressor pump | 1pc |
Suction vacuum pump | 1pc |
Main suction apparatus | 2pc |
Larynhoscope preheater | 2pc |
Monitor (LED) | 1pc |
Printer | 1pc |
Applications:
It is applicable to ENT section checking and treating patients in every hospital.
The environment and condition of usage:
Temperature: 10℃~35℃
Humidity: ≤80%
Pressure: 86kpa~106kpa
Working condition:
Power supply: 220V±10% ;50 Hz -60 Hz;
Consumption: 1200VA
Duration: continuous running time ≥8h
Safety standard:
The apparatus accords with Chinese national standard GB9706.1—1995. The safe classify of lightning strike is ClassⅠ,Type B. the apparatus accords with GB/T14710-2009.
Specifications:
a. Positive pressure of spray device: 0.1Mpa~0.15Mpa (Adjustable)
b. Negative pressure of suction device: ≥0.07Mpa (Adjustable)
c. Image processing function
Dynamic video process:
For dynamic image, real-time to freeze, capture and thaw. Real-time to display full screen . Adjust brightness, contrast, saturation.
Static image process:
For static image , partially zoom image, rotate in 90 degree, mirror left and right, mirror up and down ,negative picture . For static image ,coded false color, linearity, exponent, logarithm , enhance display. It also can note text on the picture and measure line, circle and area.
Check and print:
Various characters input methods of windows. Patient file record, delete, query and so on. Report preview and printing.