Thermal Abuse Test Chamber for Battery Module According to UL 1642 IEC62133 IEC62619
What is Thermal Abuse Test Chamber ?
Thermal abuse test chamber protect for short circuit caused by incorrect operation of the operator and battery electrode, battery positive and negative connection, etc
What are the standards for thermal abuse test chambers?
What are the relevant standard requirements ?
Heating test steps:
–For lithium storage battery, the oven temperature is raised at a rate of 5℃/min to the temperature of 130℃±2℃, remain this temperature for 30min, then stop heating.
–For metal cyanide nickel battery, the oven temperature is raised at a rate of 5℃/min to a temperature of 85℃±2℃, remain this temperature for 2h, then stop heating.
-For lithium storage battery, the oven temperature is raised at a rate of 5℃/min to a temperature of 130℃±2℃, remain this temperature for 30min, then stop heating.
–For metal cyanide nickel battery, the oven temperature is raised at a rate of 5℃/min to a temperature of 85℃±2℃, remain this temperature for 2h, then stop heating.
Thermal Abuse Test Chamber Technical Parameters
Model
|
BE-101-1000A |
Temp. Range
|
Ambient – 200℃ (Controllable) |
Control Accuracy
|
±0.5℃ |
Temp. Uniformity
|
±2.0℃( No load) |
Temp. Rising Speed
|
RT – 150℃ (≧5℃/min, liner with load 40kg module) |
Temp. overshoot
|
≦2.0℃ |
Control Display Method
|
PLC Touch screen + remote control, networking, |
Cell phone APP alarm device, with USB interface, R485 interface can be connected to computer control. | |
Testing Room Dimension
|
W1000 X D1000 X H1000 mm |
External Chamber Dimension
|
W1400 X D1300 X H1840 mm |
Inner chamber Material
|
SUS304 Stainless Steel (Mirror finished) |
External chamber Material
|
SECC Steel Plate, Powder Painted |
Basement
|
With universal wheel |
Observation Window
|
390 X 360mm (20mm Tempered Explosion Proof Glass) |
Power Source
|
380V 50HZ |
Heating Power
|
About 12 KW |
Safety Protection Device
|
over-voltage, over-temperature, over-current protection.
|
protection for short circuits caused by incorrect operation of the operator and battery electrode, battery positive and negative connection, etc. | |
Safety Device
|
Smoke exhaust system, explosion-proof pressure relief port, tri-color light alarm. |