IEC60068-2-2 Programmable Constant Temperature and Humidity Test Chamber 1000L
Complied Standards & Clauses
IEC60068-2-1:2007 “Environmental Testing-Part 2-1: Tests-Test A: Cold, IDT”
IEC60068-2-2:2007 “Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT”
IEC60068-2-78:2012 “Environmental Testing-Part 2-78: Tests-Test Cab: Damp heat, steady state, IDT”
IEC60068-2-30:2005 “Environmental Testing-Part 2-30: Tests-Test Db: Damp heat, cyclic (12h+12h cycle), IDT”
Product Overview
This test equipment prohibits:
1.Testing or storage of flammable, explosive and volatile substance samples
2.Testing or storage of corrosive substance samples
3.Testing or storage of biological samples
4.Testing or storage of samples from strong electromagnetic emission sources
Test environment conditions: When the ambient temperature is +25℃, relative humidity ≤85%, and there is no sample in the test chamber
Technical Parameters
Product name | Programmable constant temperature and humidity test chamber |
Product model | SN881-1000L |
Volume, dimensions and weight | |
Nominal internal volume | 1000L |
Internal dimensions | W1000* H1000*D1000mm |
External dimensions | W1500* H2150*D1520mm |
Weight | About 560kg |
Performance | |
Temperature range | -70℃→+150℃ |
Humidity range | (20~98)%RH (Temperature: 20℃~85℃, refer to the temperature and humidity controllable range chart, no active heat load) |
Temperature fluctuation | ±0.5℃ |
Humidity fluctuation | ≤2.5% |
Temperature deviation | ≤±2.0℃ |
Humidity deviation | ≤±3% |
Heating time | No load average 2~3℃/min |
Cooling time | No load average 0.7~1℃/min |
Load conditions | none |
Work noise | A sound level≤70dB(A) (Measured in a soundproof room with an ambient temperature of 25°C and low echo; A weighting is used to test the average of 9 points; each test point is 1 meter away from the noise source horizontally and 1 meter above the ground.) |
Structure | |
Insulated building envelope
| ①Outer wall material: A3 board spraying ②Inner wall material: Brushed stainless steel plate SUS﹟304. ③Box insulation material: 100mm high temperature resistant rigid polyurethane foam. ④Door insulation material: 100mm high temperature resistant rigid polyurethane foam. |
Bottom structural strength | The heavy capacity of the track at the bottom of the test chamber: ≤100Kg/m² (load) |
Air conditioning channel | ① Stainless steel long-axis centrifugal fans: 4 units/90W. ②Fan, heater, evaporator (also dehumidifier), drainage device, pressure balance port, adjustable air guide plate, temperature sensor |
Standard configuration of test chamber | ①Window 400x600x40mm 3-layer vacuum tempered glass. ②Flat embedded handle, ③Door hinge: SUS #304 imported hinge ④ Energy-saving lamps in the box: LED energy-saving lamps ⑤Lead hole: φ50mm 1 piece |
Box door | ①The single-piece door opens outwards, with the hinge on the left and the handle on the right (when facing the front of the box). ② Equipped with a safety door locking mechanism (the door can be opened in the test room), the door is equipped with electric heat to prevent condensation, and the viewing range of the insulating glass observation window is approximately (W400×H600mm). Electric heating device to prevent condensation on door frame |
Control Panel | The controller is a 600 7-inch display screen, temperature (humidity) control display screen, operating indicator light, and lighting button |
Mechanical room | Refrigeration unit, compressor water tray, pressure relief device, heating device |
Distribution Cabinet | Distribution board, cooling fan, main power switch, transformer, intermediate relay, time relay, solid state relay, AC contactor, circuit breaker |
Heater | ① Fin-type heat pipe-shaped stainless steel electric heater ②Heating control method: SSR (Solid State Relay) contactless equal period pulse width modulation ③Heating power: about 4KW |
Power cord hole | Located on the back of the box |
Cooling System | |
Hot and cold equipment | ① Hitachi or Emerson low-temperature and low-noise vortex ultra-low temperature refrigeration system (air-cooled heat dissipation method). ②Cooling method: air cooling |
Hot and cold exchange device | Two-stage refrigeration system Hitachi or Emerson low-noise scroll compressor. Adopting ultra-high efficiency SWEP cold coal heat and cold exchange design [made in Sweden (sus#316)], it is more efficient than the traditional internal spiral type. |
Heating load regulation | ①Adopt automatic adjustment of cold coal flow device to effectively take away the heating load of the product to be tested and German/Japanese technology Technological synchronization. ②Compared with the traditional design, the control stability and reproducibility are higher, and it achieves super efficiency of saving power and saving more energy. |
Refrigeration Compressors | Hitachi or Emerson |
High efficiency components | ①The condenser and evaporator adopt AC&R double spoiler type aluminum fin device. ②The throttling devices adopt imported Chiming refrigeration accessories. |
Expansion system | Capacity controlled refrigeration system (capillary tube). |
Evaporative condenser | All copper tube heat exchanger. |
Refrigerator control method | ①The controller of the control system automatically adjusts the operating conditions of the refrigerator according to the test conditions. ②Evaporation pressure regulating valve. ③Compressor return air cooling circuit ④ The refrigeration system is designed with vibration and noise reduction measures such as shock absorber tubes. |
Refrigerant | Honeywell R404a (ozone depletion index is 0) |
Electrical control system | |
Controller model | 600 Touch Intelligent Programmable Temperature Controller |
Controller specifications | Technical indicators: 1. 7-inch true color touch screen 2. Two control methods: program/set value 3. Sensor type: two PT100 inputs (electronic sensor input optional) 4. Output mode: voltage pulse (SSR)/control output: 2 channels (temperature/humidity)/2 channels 4-20mA analog output/16 channels relay output (passive) 5. Control signals: 8 channels IS control signals/8 channels T control signals/4 channels AL control signals 6. Alarm signal: 16 DI external obstacle alarms 7. Temperature measurement range: -90.00℃--200.00℃, (optional -90.00℃--300.00℃) error ±0.2℃; 8. Humidity measurement range: 1.0%--100%RH, error ±1%RH; 9. Communication interface: (RS232/RS485, the longest communication distance is 1.2km [optical fiber can reach 30km]); 10. Interface language type: English 11. Equipped with Chinese character input function; 12. Can bring a printer (USB function optional); 13. Multiple signal combination relay outputs, the signals can be subjected to logical operations (NOT, AND, OR, NOR, XOR), Referred to as PLC programming ability; 14. Diversified relay control methods: parameter->relay mode, relay->parameter mode, logic combination mode, composite signal mode; 15. Program editing: 120 groups of programs can be programmed, and each group of programs can have a maximum of 100 segments; |
Controller technical parameters | Accuracy: temperature ±0.1℃+1digit, humidity ±1%R.H+1digit Resolution: temperature ±0.01℃, humidity ±0.1%R.H. Temperature slope: 0.1~9.9 can be set. It has upper and lower limit standby and alarm functions. Temperature and humidity input signal dry and wet bulb PT100x2. 9 sets of P.I.D control parameter settings, P.I.D automatic calculation. Dry and wet bulb automatic correction screen |
Control method | ①Anti-integral saturation PID ②BTC balanced temperature control method + DCC intelligent cooling capacity control + DEC intelligent electrical control (temperature test equipment) ③BTHC balanced temperature and humidity control method + DCC intelligent cooling capacity control + DEC intelligent electrical control (temperature and humidity test equipment) |
Screen display function | 1. It adopts screen conversation style, no key input is required, and the options can be directly touched on the screen. 2. Temperature set (SV) and actual (PV) values are directly displayed. 3. It can display the currently executing program number, segment, remaining time and number of cycles. 4. Operation accumulated time function. 5. The temperature program setting value is displayed in a graphic curve, with the function of real-time display of program curve execution. 6. It has a separate program editing screen, and at least 5 segments of temperature, humidity and time can be input on each page. 7. Chinese and English can be switched at will. 8. The fault prompt screen is displayed. 9. The screen backlight can be adjusted. 10. The screen display protection function can be set to timer, TIMER or manual shutdown. |
Program capacity and control functions | 1. Usable program groups: maximum 120 PATTEN. 2. Usable memory capacity: 12000 SEGMENTS in total. 3. Repeatable execution of commands: each command can be executed up to 3200 times. 4. The program is created in a conversational style, with functions such as editing, clearing, and inserting. 5. SEGMENTS time setting 0~99Hour59Min. 6. Programmable timing control module device x2 set. 7. It has a power-off program memory, and automatically starts and resumes program execution after power is restored. 8. With RS-485 or RS-232 communication interface. 9. Graphical curves can be displayed in real time when the program is executed. 10. It has the function of automatically adjusting the freezing capacity. 11. It has scheduled start-up and shutdown functions. 12. With date and time adjustment function. 13. Button and screen lock (LOCK) function. |
Safety protection device | |
Cooling System | Compressor overheating Compressor overflow Compressor overpressure |
Test chamber | Adjustable over-temperature protection device Air conditioning channel extreme overtemperature Fan motor overheated |
Other | Leakage Protection Heating tube no air drying protection Overload and short circuit protection Power outage protection low water level protection |
Other configurations | |
Sample main power leakage circuit breaker | AC380V, 50Hz, three-phase four-wire + protective ground wire. |
Other special instructions | |
Power cable | 5-core (three-phase four-wire + protective ground wire) cable 4 meters, 1 piece (can be provided according to customer requirements) |
Lead hole | 1 lead hole with rubber plug, diameter φ50mm, its location and quantity can be customized according to user requirements as long as the box structure allows and does not affect performance. |
Transportation | Car packaging and transportation |
Conditions of Use | List the conditions |
Installation site | Flat ground and good ventilation No strong vibration around the device There is no strong electromagnetic field influence around the device There are no flammable, explosive, corrosive substances and dust around the equipment Leave appropriate space for use and maintenance around the equipment, as shown in the figure below: AB distance is about 1 meter
|
Environmental conditions | ① Temperature: 5℃~35℃ ②Relative humidity: ≤85% ③Air pressure: 86kPa~106kPa |
Power supply conditions | AC380V three-phase four-wire + protective ground wire |
Power supply | Allowable voltage fluctuation range: AC (380±10) V Allowable frequency fluctuation range: (50±0.5) Hz The grounding resistance of the protective ground wire is less than 4Ω The user is required to configure an air or power switch of corresponding capacity for the equipment at the installation site, and this switch must be independent for use by the equipment. |
Battery capacity | About 11kW |
Maximum current | 30A |
Switch | 63A (molded case leakage protector) |
Requirements for storage environment | When the equipment is not working, the ambient temperature should be kept within 0℃~+45℃ |