IEC60068-2-30 Walk-in Constant Temperature -and Humidity Test Chamber 6.8m³
Complied Standards & Clauses
Conforms to:
IEC60068-2-1:2007 “Environmental Testing-Part 2-1: Tests-Test A: Cold, IDT”
IEC60068-2-2:2007 “Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT”
IEC60068-2-78:2012 “Environmental Testing-Part 2-78: Tests-Test Cab: Damp heat, steady state, IDT”
IEC60068-2-30:2005 “Environmental Testing-Part 2-30: Tests-Test Db: Damp heat, cyclic (12h+12h cycle), IDT”
Product Overview
The walk-in constant temperature and humidity test room is suitable for safety performance testing, reliability testing, product screening testing, etc. of electronic components. At the same time, through these tests, the reliability of the product is improved and the quality control of the product is carried out.
The walk-in constant temperature and humidity test room is an essential test equipment in the fields of aviation, automobiles, home appliances, scientific research, etc. Sample range: electrical, electronic, semiconductor, communication, optoelectronics, electrical appliances, automotive electrical appliances, materials and other products. This equipment is used to assess and determine the parameters, performance, and adaptability of the above samples after the high and low temperature and humidity tests with rapid changes in temperature environment.
It is suitable for schools, factories, military industry, research and development and other units.
Technical Parameters
1.1 Model and Purpose | |
1.1.1 Product model | Walk-in Constant Temperature And Humidity Test Chamber SN886-6.8m³ |
1.1.2 Purpose | The walk-in constant temperature and humidity test room is suitable for safety performance testing, reliability testing, product screening testing, etc. of electronic components. At the same time, through these tests, the reliability of the product is improved and the quality control of the product is carried out. The walk-in constant temperature and humidity test room is an essential test equipment in the fields of aviation, automobiles, home appliances, scientific research, etc. Sample range: electrical, electronic, semiconductor, communication, optoelectronics, electrical appliances, automotive electrical appliances, materials and other products. This equipment is used to assess and determine the parameters, performance, and adaptability of the above samples after the high and low temperature and humidity tests with rapid changes in temperature environment. It is suitable for schools, factories, military industry, research and development and other units. |
1.2 Volume and Size | |
1.2.1 Effective volume | 6.8m3 |
1.2.2 Inner box dimensions | W: 2000 mm H: 1800 mm D: 1900 mm |
1.2.3 External dimensions Approx. | W: 2200 mm H: 2200 mm D: 3200 mm |
1.3 Performance (ambient temperature -5℃~+28℃, no-load condition) | ||
1.3.1 Temperature range | -50℃~+180℃ | |
1.3.2 Heating rate | about 2℃/min (no load, average) | |
1.3.3 Cooling rate | about 2℃/min (no load, average) | |
1.3.4 Resolution accuracy | 0.01℃ | Data measured under no-load conditions |
1.3.5 Temperature fluctuation | ±0.5℃ | |
1.3.6 Temperature deviation | ±2.0℃ | |
1.3.7 Humidity range | 20%~98%RH | |
1.3.8 Humidity deviation | Below 75%RH: ≤±5.0%RH Above 75%RH: ≤+2/-3%R.H | |
1.3.9 Humidity fluctuation | ≤±2.0%R.H | |
1.3.10 Temperature and humidity control capability range diagram | ![]() | |
1.3.11 Power supply | AC3¢5W 380V 50Hz (R, STN three-phase five-wire plus ground wire) (voltage fluctuation ≤ ±10%) | |
1.3.12 Power | The total power of the equipment is about 25KW | |
1.3.13 Noise | Measured according to GB/T14623-2008, the noise level is ≤75dB (the noise detection device is measured at a distance of 1m from the equipment door). | |
1.3.14 Standards | IEC60068-2-1:2007 “Environmental Testing-Part 2-1: Tests-Test A: Cold, IDT” IEC60068-2-2:2007 “Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT” IEC60068-2-78:2012 “Environmental Testing-Part 2-78: Tests-Test Cab: Damp heat, steady state, IDT” IEC60068-2-30:2005 “Environmental Testing-Part 2-30: Tests-Test Db: Damp heat, cyclic (12h+12h cycle), IDT” | |
1.3.15 Energy-saving design | The PLC (Programmable Logic Controller) of the control system is used. The operating conditions of the refrigerator are automatically adjusted according to the test conditions. The refrigeration system can adjust the power according to the load size. It also has compressor energy regulation technology. It can achieve excellent energy-saving effects. | |
1.3.16 Cooling method | The equipment adopts water cooling (Water tower, water pump) |
Details