Battery Explosion-proof Test Chamber 1092L EV Battery Testing Equipment
Complied Standards & Clauses
Conforms to IEC60068-2-2:2007, IEC60068-2-1:2007, etc.
Equipment Overview
Environmental conditions:
1. Temperature: 5℃~35℃
2. Relative humidity: ≤85%
3. Air pressure: 86kPa~106kPa
Limitation of specimen:
This test equipment prohibits:
1. Testing or storing samples of flammable, explosive, and volatile substances.
2. Testing or storing samples of corrosive substances.
3. Testing or storing biological samples.
4. Testing or storing samples of strong electromagnetic radiation sources.
Technical Parameters
Technical Parameters | ||
Control & Operation | PLC and touch screen intelligent control system | |
Chamber Size
| Inner Size | External Size |
W1300*700H*D1200mm | W1875*1980H*1723 mm | |
Temperature Range | -70 °C ~+150°C | |
Temperature fluctuation | ±0.5℃ | |
Humidity fluctuation | ≤2.5% | |
Temperature uniformity | ≤±2.0℃ | |
Temperature deviation | ≤±1.0℃ | |
Heating Rate | -40℃→+72℃ No-load Linearity Within 30℃/min | |
Cooling Rate | +72℃→-40℃ No-load Linearity Within 30℃/min | |
Load condition | None | |
Working noise | A sound level≤70dB(A) (Measured in a soundproof room with an ambient temperature of 25°C and low echo. Using A weighting, test the average of 9 points. Each test point is 1 meter away from the noise source horizontally and 1 meter above the ground.) |
Explosion-proof pressure relief vent | A set of explosion-proof pressure relief vents is set on the left side of the equipment, with a size of 300mm×300mm. When the internal pressure of the box is ≥0.104MPa, the pressure relief vent will open automatically
|
Explosion-proof chain |
|
Explosion-proof window |
|
Control panel | Controller 7-inch display screen, temperature (humidity) control display screen, operating indicator light, and lighting button. |
Mechanical room | Refrigeration unit, compressor water tray, pressure relief device, heating device |
Power distribution cabinet | Distribution board, cooling fan, main power switch, transformer, intermediate relay, time relay, solid-state relay, AC contactor, circuit breaker |
Heater | 1. Fin-type radiator-shaped stainless steel electric heater 2. Heating control method: SSR (Solid State Relay) non-contact equal period pulse width modulation 3. Heating power: approximately 3KW |
Power cable hole | Located on the back of the chamber |