UL94 Horizontal and Vertical Flammability Testing Equipment 500W 50W
Complied Standards & Clauses
IEC 60695-11-10 “Fire hazard testing -- Part 11-10: Test flames -- 50 W horizontal and vertical flame test methods”
IEC 60695-11-20 “Fire hazard testing - Part 11-20: Test flames - 500 W flame test method”
IEC 60695-11-4 “Fire hazard testing - Part 11-4: Test flames - 50 W flame - Apparatus and confirmational test method”
UL 94 “Tests for Flammability of Plastic Materials for Parts in Devices and Appliances”
IEC 60695-11-3 “Fire hazard testing - Part 11-3: Test flames - 500 W flames - Apparatus and confirmational test methods”
Equipment Overview
Application: It simulates the effect of initial flames when a fire occurs near electrical and electronic devices, helping to assess the ignition risk. This test is primarily used to examine the combustibility of plastics and other non-metallic substances, solid materials, and foam plastics (with a density of at least 250kg/m³) following the ISO 845 testing method. The test is carried out in both horizontal and vertical orientations..
Test Principle: By controlling the gas flow and the duration of the flame exposure, the heat intensity applied to the sample can be regulated, allowing for the simulation and assessment of the fire hazard level.
Structure: This series of flammability test chambers features an integrated air intake automatic ignition system with a combustion delay of just 0.1 seconds, ensuring complete combustion of the gas. The chamber is constructed from steel, with a black matte finish on the interior for better visibility. It is equipped with a large observation window and an imported high-precision digital display, ensuring accurate monitoring and detection.
Technical Parameters
Gas Flow | 50W:105±5ml/min or 18±1 mm; 500W:965±30ml/min or 117±3 mm,adjusted by flowmeter | |
Control and Operation Mode | PLC control, operated on 7-inch color touch screen, with infrared remote control | |
Burner Standard | IEC60695-11-3/IEC60695-11-4/10,UL94,inner diameter of Φ9.5±0.3mm, length100mm,made of brass | |
Gas Pressure | 0.1MPa | |
Lighting | Equipped with a light, with control | |
Test Chamber and Dimensions | >0.75 cubic, black matte background, with exhaust fan, dimensions: 1100*550*1200mm, 140kg | |
After Flame / afterglow Time | 0-9999X0.1S can be displayed | |
Burning Time | 0-9999X0.1S, can be preset | |
Number of Flame Applications | 0-9999 times,can be preset | |
Test Flame Calibration | Distance between the bunsen burner port to temperature measuring copper block:50W: 10±1 mm; 500W: 55±1 mm | |
Thermocouple: The American original Omega φ0.5 armour nickel-chromium/nickel aluminum wire K-Thermocouple, temperature resistance 1100° | ||
Temperature rise time from100±2°to 700°±3°: 50W: 44±2S 500W:54±2S | ||
Flame height:50W: 20±2mm(there is no internal and external flame); 500W: Inner flame:40±2mm,Outer flame:125±10mm | ||
Temperature measuring copper head:50W:Ф5.5mm±0.01mm,weight before drilling 1.76 ± 0 .01 g;500W:Ф9mm±0.01mm,weight before drilling 10 ± 0 .05g | ||
Back Pressure Water Column Height of U-Shaped Pressure Gauge | 50W: 10mm 500W:125±5 mm | |
Test Angle | 50W: 0°,45° 500W: 20° |