Cosmetics Testing Equipment Melting Point Apparatus
Product introduction
The melting point meter adopts photoelectric automatic detection, dot-matrix graphic LCD display, touch screen buttons and other technologies, with melting curve, initial melting, final melting display and other functions. Its temperature control system uses precision platinum resistors as temperature detection elements, and uses digital PID adjustment and PWM temperature control technology to improve the accuracy and reliability of melting point determination.
Technical parameters
Measuring range room temperature: -400℃
Measurement method : Automatic
Minimum indication: 0.1℃
The heating rate can be continuously selected between 0.1℃/min-20℃/min
Indication error: ≤200℃ ±0.4℃ >200℃ ±0.7℃
Repeatability of indication, heating rate :1℃/min, 0.3℃
Linear heating rate error :±10%
Heating time :50℃-400℃≤5min 400℃-50℃≤7min
Pre-stored measurement mode ————
Processing capacity :1
Capillary size outer diameter: φ1.4mm inner diameter φ1.0mm length 90mm
Sample height: 3-5mm
Data storage: 1000 groups
Printer model ————
Display mode: 5.7 inch LCD screen
Calibration method Multi-point calibration
Communication interface: RS232/USB
Power supply: 220V±22V, 50 Hz
Instrument size: 360mm×290mm×170mm
Net weight of the instrument : 9kg
Product features
1. The instrument has the function of automatically saving working parameters and storing measurement results
2. The connection can be established with the PC through the communication interface.
3. The instrument uses the capillary specified in the Pharmacopoeia as the sample tube.