Needle Flame Tester IEC 60695 Flame Test Machine Testing Apparatus
Instrument
The Needle Flame Tester is a fully automatic instrument contained in its own cabinet to maximize the safety of the operator.
The Needle Flame Tester consists of a diameter of 0.9mm needle burner that tilts to 45degree from the vertical and is fueled with butane gas. The test time for the temperature of the standard copper block to increase from 100 degree to 700 degree shall be 23.5s. Test flame height is adjusted 12mm.The fire hazard of the specimen is assessed by measuring the burning duration of the specimen, and any ignition of the wrapping tissue and white pine board below the specimen. The instrument is designed to conform to IEC60695-11-5, IEC60695-2-2 and other most commonly used standards in the market.
Application
Insulation and combustible materials used inside electrical and electronic products may propagate flames when a failed component ignites, which may develop into a fire that endangers the safety of operators and product operation. The Needle Flame Tester is used to assess the fire hazard of such component materials. The test is performed to determine that the test flame does not cause ignition of parts, or that a combustible part ignited by the test flame has a limited duration of burning or a limited extent of burning, without spreading fire by flames or burning or glowing particles falling from the test specimen.
Conforms To
IEC60695-11-5, IEC60695-2-2, IEC60745-1, IEC60335-1,
IEC60950-1, IEC60598-1, IEC60065
Specification:
1, Needle burner: the hole Φ 0.5mm ± 0.1mm, outer diameter of ≤Φ 0.9mm, length ≥ 35mm, stainless steel
2, Burner angle: tilted 45 ° (when testing) or vertically (when adjust the flame height)
3, Burning time: 0-999.9s ± 0.1s adjustable (usually selected 30s)
4, Burning time: 0-999.9s ± 0.1s, auto record, manual pause
5, Flame height: 12mm ± 1mm (with height measuring apparatus)
6, Burning gas: 95% butane gas (lighter gas can be bottled)
7, Temperature measuring range: 0 ~ 1000 ℃
8, the flame temperature requirements: the time to rise to 700 ℃ ± 3 ℃ from 100 ℃ ± 2 ℃ in 23.5 seconds ± 1 seconds
9, thermocouple: Φ0.5mm imported armored (K type) thermocouple
10. Test Background: Black background
11, the testing process: test procedures automatic control, independent ventilation
12, reference standard: IEC60695 -11-5, GB / T5169.5-2008, GB4706.1-2005
13,Dimensions: 1100mmX700mmX1300mm
Needle Burner | the hole Φ 0.5mm ± 0.1mm, outside diameter ≤ Φ 0.9mm, length ≥ 35mm, |
Main Material | SUS304 stainless steel |
Needle Flame Burner Angle | tilt 45 ° (test) or vertical (adjust the flame height) |
Application Time | 0-999 .9 s ± 0.1s adjustable (normally choose to 30s) |
Burning Duration | 0-999 .9 s ± 0.1s, auto record, manual suspension |
Needle Flame Height | 12mm ± 1mm (with a height measurement apparatus) |
Burn Gas | 95% of the butane gas (Pressure:Less than 0.2MP) |
Temperature Measurement range | 0 ~ 1000 ℃ |
Needle Flame Temperature Requirements | from 100 ℃ ± 2 ℃ 700 ℃ ± 3 ℃ rise to the time in 23.5 seconds ± 1 seconds |
Thermocouple | Φ0.5mm (K type) |
Experimental Background | black background |
Test Procedure | Button/control panel |
Reference Standard | IEC60695 -11-5-2004,UL746A,GB5169.5-2008,GB4706.1 |
Inside Dimension | 0.5 cubic meters, stainless steel chassis |
Equipment Dimensions | About 1000W× 650D × 1300H mm |
Power Supply | AC220V ± 5%, 50HZ |
Our service
Pre-sale service:
1. One-on-one pre-sales consulting service to answer your questions
2. Recommend the right products for your needs
3. Visit factory support
After-sale service:
1. All of our products have one year free warranty.
2. We have a professional after-sale solution team to solve your probelm
3. Provide technology support
4. If necessary, home repair