Computer Closed Flash Point Flammability Test Chamber With Color LCD Display
Scope of application:
It is applicable to the determination of oil product closing flash point value and widely used in power, petroleum, chemical industry, commodity inspection, scientific research and other departments.
Meet the standard:
Standard method for ASTM D93, GB/T 261-2008
Main features:
3.1 adopting ARM microprocessor technology, color LCD, resistive touch screen technology, Chinese menu, human-computer interaction is more convenient;
3.2 the power storage function of the instrument;
3.3 the device has the functions of automatic ignition, display, locking and printing results, automatic cooling, etc.
3.4 atmospheric pressure automatic correction. The instrument has the function of the external adjusting function of igniter brightness.
The instrument has the advantages of accurate measurement, good repeatability, stable performance and simple operation.
Main parameters:
4.1 display device: color LCD display
4.2 operation mode: touch screen
Measuring range: 4.3 at room temperature to 350 ℃
4.4 temperature detection: platinum resistance
4.5 weight after sex: > 110 ℃ + 2 ℃ 110 ℃ or less + / - 1 ℃
4.6 ignition mode: electric ignition
4.7 information storage: can store 500 analysis results
4.8 cooling mode: forced air cooling
4.9 Printer: thermal printer
4.10 recomplex: conforming to ASTM D93 GB/T 261 standard
4.11 Source: ac 220V plus or minus 11V, 50Hz plus or minus 2.5hz
4.12 work rate: no more than 350VA
4.13 using the environment temperature: 10 ℃ ~ 35 ℃
4.14 Ambient humidity: less than 85%