High Accuracy Children's Forehead Thermometer Fast Temperature Measurement
Product Description
1. Non Contact Infrared Body Thermometer, specially designed for testing baby temperature, best baby thermometer
2. Multiple Function: 0-100℃(32-212℉) wide range, not only for body temperature, but also for temperature of surface of the item like milk, food, bath water etc
3. 5-15cm(2-5.9in) measuring range: Safe and clean
4. Quick and accurate: In just 1seconds with ±0.1℃/℉ precision
5. ℃ and ℉ switchable
6. 32 memories
8. Setting be adjustable for different kinds of people and operating conditions
9. Automatic shut off: 15 seconds without operation
Specification
S/N
|
Feature
|
Description
|
1
|
Product name
|
Baby Infrared Forehead Thermometer
|
3
|
Size:
|
153*41*44mm
|
4
|
Main Features
|
32 reading memory, dual scale, cpmpact ergonomic size, beeps when ready, jumbo display, replaceable battery, auto-off, 3 seconds reading
|
5
|
Measuring site
|
Forehead(Forehead Mode)
|
6
|
Measuring range
|
Forehead mode: 34.0℃~43.0℃(93.2℉~109.4℉)
Object mode: 0℃~100℃(32℉~212℉) Room temperature: 0℃~50℃(32℉~122℉) |
7
|
Reference body site:
|
Oral (This thermometer converts the ear temperature to display its "oral equivalent.")
|
8
|
Operation mode:
|
Forehead mode(Adjust mode)
|
9
|
Laboratory accuracy
|
Forehead mode:
±0.2℃ (0.4℉) during 35.5℃~42.0℃ (95.9℉~107.6℉) at 15℃~35℃ (59.0℉~95.0℉) operating temperature range ±0.3℃ (0.5℉) for other measuring and operating temperature range Object mode: ±4% or ±2℃(4℉) whichever is greater Room temperature: ±2℃(±4℉) |
10
|
Display resolution
|
0.1℃ or 0.1℉
|
11
|
Measure time
|
Approximately 3 second
|
12
|
Operating temperature
range: |
Forehead/Object mode:
10℃~40℃(50℉~104℉), 15%~85%RH, non-condensing Atmospheric Pressure : 700hPa ~ 1060hPa |
13
|
Storage and transport
temperature range |
-25℃~ 55℃ (-13℉~131℉), 15%~95%RH, non-condensing Atmospheric Pressure : 700hPa ~ 1060hPa
|
14
|
Warranty
|
2 year
|
15
|
Clinical accuracy
|
Clinical bias: -0.06℃(-0.1℉)
Clinical repeatability: 0.13℃(0.23℉) Limits of agreement: 0.85℃(1.53℉) |
Measurement considerations
FAQ
Q: How to switch thermometer unit symbols between ℃ and ℉ ? .
A: When the thermometer is turned off, press and hold the scan key until the temperature unit symbol (℃ or ℉) appears on the screen,then you can switch thermometer unit symbols between ℃ and ℉ through click the scan key now..
Q: Need I debug the forehead thermometer before use?
A: No, the thermometer passed a rigorous calibration test before it leaves the factory, and the thermometer can automatically adapt to the ambient temperature and automatically debug it..
Q: What to do if I have a problem during using it.
A: Whenever you meet a problem during using our problem,you can ask our Customer Service for help.