Specifications
Brand Name :
Teledyne
Model Number :
8800A
Certification :
ROHS
Place of Origin :
Teledyne
MOQ :
1PC
Price :
negotiable
Payment Terms :
T/T
Supply Ability :
1~1000pcs
Delivery Time :
on request
Packaging Details :
carton
Origin :
USA
Warranty :
1 year
Manufacturer :
Teledyne
Products Name :
Teledyne Trace Moisture Analyzer
Model :
8800A
response time :
45 seconds
​Range by sensor type: :
XTR-100 -148° to 68°F (-100° to 20°C)
Pressure Operating Range :
​Range by sensor type:
​Accuracy :
​± 5.5°F (± 3°C)
​Repeatability :
​± 0.9°F (± 0.5°C)
Description

Teledyne Trace Moisture Analyzer​​ 8800A



Part Number: 8800A
Product Name: Trace Moisture Analyzer
Origin: USA
Manufacturer: Teledyne


Specification:

The 8800 series uses field proven aluminum oxide (Al2O3) sensing technology to accurately detect trace moisture on either a continuous or spot checking basis. All Al2O3 sensors share the same basic operating principle: the capacitance measured between the sensor’s aluminum core and gold film deposited on the oxide layer varies with the water content. The 8800 series moisture sensor employs unique Hyper Thin Film (HTF) technology, which offers three major structural improvements in Al2O3 sensor design. These structural changes, noted below, provide the user with increased sensitivity, greater stability and a quicker response time when compared to other conventional aluminum oxide sensors on the market today.




Feartures

Readout available in (5) engineering units; °C, °F, ppm, lbs. H2O/mm scf & g H2O/m3

Scalable signal outputs (0-20 mADC & 4-20 mADC)

Built-in LCD with backlight, 3.5 digits and custom legends for units / mode

Din box enclosure: (cut-out dimensions of 137 mm W x 67 mm H)

(4) front panel membrane tactile push button controls, all settings stored in EEPROM

(2) programmable alarm relays w/programmable variable hysteresis rated 10A @ 240V

Power requirements of 100-240 VAC, 50 or 60 Hz

Operating temperature of electronics: -10 to 50° C

Class I & II, Division 1, Groups A, B, C, D, E, F & G (when used with optional safety barriers)

High capacitance aluminum oxide (Al2O3) sensor using

Hyper Thin Film (HTF)™ Technology

(2) meters of sensor interconnection cable included. See Price-Adders for longer cable

Power cord

Instruction manual (one copy per analyzer ordered)​




A-1: Obsolete, contact factory
A-2C
oxygen sensor
C6689-A2C Used for trace O2 measurements in CO2 containing backgrounds (0.1-100% CO2). CO2 must be present. Use where acid gases are present (CO2, HCl, HCN, etc). CO2 must be present.
A-2CXL
oxygen sensor
B74033-A2CXL Specially qualified A2C sensor to operate in XL model analyzers. Low range limit is 0-1 ppm. Use where acid gases are present (CO2, HCl, HCN, etc). CO2 must be present.
A-2Cl: Obsolete, use A-2C
A-3: Obsolete, use A-5
A-5
oxygen sensor
C6689-A5 Intermittent or continuous monitoring of percent O2 in gas mixtures containing 0-100% CO2. 90% response is time is 45 seconds. Has more tolerance to exposure to air than A-3. No restrictions on CO2 concentration.
B-1
oxygen sensor
C6689-B1 General purpose percent oxygen analysis where fast response is desired (90% in less than 7 seconds). Use B-3 or C-3 where longer life is required and response time is relatively unimportant. May be used to spot check flue gas.
B-2C
oxygen sensor
C6689-B2C General purpose trace O2 analysis for use in inert gas and hydrocarbon streams. Low range limit is 0-10 ppm. May be used in CO2 free gas streams containing highly mobile gases such as H2, He and ethylene.
INSTA TRACE
oxygen sensor
B71875 Special patented construction provides rapid recovery, during initial cell installation, to low PPM levels by reducing the recovery period from several hours to just minutes. Low range limit is 0-10 ppm. Used for trace O2 analysis in inert gas and hydrocarbon streams.
B-2CXL
oxygen sensor
C6689-B2CXL Specially qualified B-2C sensor to operate in XL model analyzers. Low range limit is 0-1 ppm. May be used in CO2 free gas streams containing highly mobile gases such as H2, He and ethylene.
B-3
oxygen sensor
C6689-B3 Percent O2 analysis, best compromise between long life and fast response. May be used to spot check flue gas.
B-4C
oxygen sensor
C14613-B4C Trace O2 cell specifically designed for the Model 317EXI. Same performance as the B-2.
B-5
oxygen sensor
C45091-B5 Special application cell (non-standard) where long life and moderately fast response are required; B-5 has a factor of 3x the life of a B-3. Requires 3x the amplifier gain of the B-3. Expected life is 36 months in air.
B-5F
oxygen sensor
C45091-B5F Special application cell (non-standard) where long life and moderately fast response, in the presence of CO2, are required. Uses Type F (acid electrolyte). Requires higher amplifier gain than normal. 90% response time is 13 seconds. Expected life in air is 6 months.
B-7
oxygen sensor
C45091-B7 Special application cell (non-standard) where fast response is required; the B-7 has a factor of 3x the life of a B-1 cell. Requires 3x the amplifier gain of the B-1. Expected life in air is 24 months.
B-7F
oxygen sensor
C45091-B7F Special application cell (non-standard) where moderate life and fast response, in the presence of CO2, are required. Requires higher amplifier gain than normal. 90% response time is 7 seconds. Expected life in air is 3 months.
C-1
oxygen sensor
C6689-C1 Rarely used. Schematic for reference only.
C-2
oxygen sensor
C6689-C2 Rarely used. Schematic for reference only.
C-3
oxygen sensor
C6689-C3 Percent O2 measurements where maximum cell life is desired (high O2 applications). 90% response time is 30 seconds. Expected life in air is 18 months.
C-5
oxygen sensor
C45091-C5 Special application cell (non-standard) where long life is required; the C-5 has a factor of 3x the life of a C-3 cell. Requires 3x the amplifier gain of the C-3. Expected life in air is 54 months.
C-5F
oxygen sensor
C45091-C5F Special application cell (non-standard) where long life, in the presence of CO2, is required. Requires higher amplifier gain than normal. 90% response time is 30 seconds. Expected life in air is 9 months.
C-6
oxygen sensor
C54090-C6 Motion insensitive percent O2 measurements for use in hand held portable monitors. Best cell for OEM safety applications.
E-1
oxygen sensor
C14991-E1 Special cell with performance characteristic of the C-3 with 5x the life. Cell is 2 to 3 times the length of standard MFCs; requires special mounting. Designed for medical oxygen concentrators.
E-2
oxygen sensor
C57283-E2 Long life cell for use is the Model 3290 (for use in unattended air liquefaction nitrogen skids). Has a 3 to 4 years expected life in air. Minimum range 0-1000 ppm.
F-1R C40371-F1R 0-1000 ppm CO in Air Used in Max 5 and 9950 Portable Combustion Analyzers

Our Advantage:

Best possible prices and delivery time

Long time experience in the field

Well trained staff

Own warehouse facilities

Best relationships with branch office of manufacture

Flexible and individual service for you

Cheap and safe forwarder for shipments (air)



Frequently Asked Questions:

Q: Does your goods are new and original?

A: Yes, they are new and original.


Q: How long warranty?

A: One-year warranty.


Q: Does any certificates can be provided?

A: Yes, it depends on what product you request, please check details with our sales.


Q: Varied payment methods.

A: We adopt T/T, alipay and paypal and Currency in EURO,USD,RMB


Q: SHIPPING:

A: We can shipping all over the world.via DHL,UPS,Fedex,TNT and IPS,PDE,BRE,KARAPOST etc.


Our Product Range:

Transmitter, Safety barries, Flow meters, Probe, Sensor, Transducer. Valves,Analysis meter, Positioner, Acuator, etc.

240V 8800A Al2O3 Teledyne Analytical Instruments Trace Moisture ​​

240V 8800A Al2O3 Teledyne Analytical Instruments Trace Moisture ​​


Send your message to this supplier
Send Now

240V 8800A Al2O3 Teledyne Analytical Instruments Trace Moisture ​​

Ask Latest Price
Brand Name :
Teledyne
Model Number :
8800A
Certification :
ROHS
Place of Origin :
Teledyne
MOQ :
1PC
Price :
negotiable
Contact Supplier
240V 8800A Al2O3 Teledyne Analytical Instruments Trace Moisture ​​
240V 8800A Al2O3 Teledyne Analytical Instruments Trace Moisture ​​

GREAT SYSTEM INDUSTRY CO. LTD

Verified Supplier
1 Years
Since 1998
Business Type :
Distributor/Wholesaler
Total Annual :
7000000-10000000
Employee Number :
50~100
Certification Level :
Verified Supplier
Contact Supplier
Submit Requirement