Features:
Ø 3.5 inch LCD screen, multiple languages customization
Ø Wide range of flow rate:0.01-2000ml/h
Ø Comprehensive audible and visual alarm function, adjustable volume, visible from 360°
Ø Automatic recognition of syringe specifications(5ml,10ml,20ml,30ml,50/60ml)
Ø Change flow rate without stopping the current task
Ø Drug library mode, customized
Model No. Parameters | DY115 | |||
Syringe Size | 5,10,20,30,50/60ml | |||
Drug Library | Yes | |||
Infusion Modes | Flow Rate,Rate-Time,Rate-Volume,Time-Volume, Body Weight,Drug Library |
|||
Drug Name Display | Default 12 drugs, can be updated & added to 2000 | |||
Safety | Double CPU to make sure higher standard infusion | |||
Accuracy | ±2% | |||
Time Preset | 0~99h and 59min | |||
Volume Range | 0.01~9999.99ml | |||
Flow Rate Step by | Increments at 0.01 ml/h | |||
Flow Rate Range | 5ml Syringe 0.01ml/h-150ml/h | |||
10ml Syringe 0.01ml/h-400ml/h | ||||
20ml Syringe 0.01ml/h-600ml/h | ||||
30ml Syringe 0.01ml/h-900ml/h | ||||
50/60ml Syringe 0.01ml/h-1800ml/h | ||||
Purge/Bolus Rate/Top Flow Rate | 5ml Syringe 150ml/h | |||
10ml Syringe 400ml/h | ||||
20ml Syringe 600ml/h | ||||
30ml Syringe 900ml/h | ||||
50/60ml Syringe 1800ml/h 50/60ml Syringe 1200ml/h |
||||
Audible & Visible Alarms | Syringe Dislocation,Occlusion,Near End,Pause Overtime,Low Battery,Battery Exausted,VTBI Completion,Syringe Empty,Motor Malfunction,Incorrect Syringe Specification,Circuit Malfunction,Internal Communication Malfunction,Master CPU Malfunction,Monitoring CPU Malfunction | |||
KVO | Rate≥10ml/h:KVO=3ml/h Rate<10ml/h:KVO=1ml/h | |||
Occlusion Pressure | High: | 106.7KPa±26.7 KPa | ||
Middle: | 66.7KPa±13.3 KPa | |||
Low: | 40.7KPa±13.3 KPa | |||
Pause Overtime Alarm | 1min50s—2min | |||
Power & Battery | 20VA, AC:100V-240V 50HZ/60HZ, DC:11.1V,Lithium battery support from 7 to10 hours | |||
Classification | Class II,type CF,IPX4 | |||
Dimension & Weight | 140mm(width)×260mm(length)×110mm(height) ; 2.5Kg | |||
Optional Function | Event Log with at least 2000 history records,DC |