BH93420-WS water level transmitter sensor fuel sensor with Flange
BH93420-WS Sewage Level SensorPerformance Parameters of water level pressure transmitter sensor | |
---|---|
Measurement Range | 0~10m…400m |
Pressure Type | Gauge Pressure, Absolute Pressure, Sealed Pressure |
Overload | ≤ 2X rated range |
Accuracy @ 25ºC | ±0.25% (Typ.) ±0.5% (Max.) |
Repeatability | ±0.03% (Typ.) ±0.05%FS (Max.) |
Hysteresis | ±0.03% (Typ.) ±0.05%FS (Max.) |
Long-Term Stability | ±0.2%F.S/year (Max.) |
Zero Temperature Drift | ±0.01%FS/ºC(>100KPa), ±0.02%FS/ºC (≤100Kpa) |
Span Temperature Drift | ±0.01%FS/ºC(>100KPa), ±0.02%FS/ºC (≤100Kpa) |
Compensation Temperature | -10-80ºC |
Operating Temperature | -20~100ºC |
Storage Temperature | -30~100ºC |
Vibration | 10g,55Hz ~ 2kHz |
Housing Material | 316L Stainless Steel |
Diaphragm Material | 316L Stainless Steel |
Insulation Resistance | >=100MΩ@100VDC |
Explosion-proof Rating | ExiaIICT6 |
Protection Rating | IP67, IP68 |
O-Ring Seal | Fluoro rubber |
Weight | ~2.1Kg (without cable) |
Cable Material | Polyethylene, Polyurethane,PTFE |
Output Signal | Supply Voltage | Output Type |
4~20mA | 12~30VDC | 2/3/4 wire |
0~10/20mA | 3 wire | |
0/1~5V | ||
0/1~10V | ||
0.5~4.5V | 5VDC |
Outline Construction |
Selection Examples |
Ordering tips |
1. When selecting, please ensure that the tested medium is compatible with the product parts in contact with it.
2. The default medium is water and the density is 1000kg/m3. If the measuring medium is other liquid or water with different density, the density of the measured medium should be provided for more accurate measurement.
3. For strong electromagnetic or radio frequency interference, please provide us with detailed information and indicate in the order.
4. In order to ensure the reliable operation of the product, it is recommended that users install lightning protection devices on site and ensure that the product and power supply are reliably grounded.
Order Guide of water level pressure transmitter sensor | ||||||||
BH | Code | Outline Construction | ||||||
WS | Pressure Transmitter | |||||||
Measurement Range | 0~10m…400m | |||||||
Code | Power Supply | |||||||
D1 | 24VDC | |||||||
D2 | 5VDC | |||||||
D3 | Other | |||||||
Code | Output Signal | |||||||
S1 | 4~20mADC | S5 | 0~20mADC | |||||
S2 | 1~5VDC | S6 | 0~10VDC | |||||
S3 | 0~5VDC | S7 | 0.5~4.5VDC | |||||
S4 | 0~10mADC | |||||||
Code | Pressure Type | |||||||
G | Gauge Pressure | |||||||
A | Absolute Pressure | |||||||
S | Sealed Reference Pressure | |||||||
Code | Cable Material | |||||||
C1 | PE | |||||||
C2 | PU | |||||||
C3 | PTFE |