With Color large-screen LCD display, touch screen buttons, fully digital parameter operation, etc., built-in reverse control chromatography workstation, realize instrument reverse control, with high integration, simple operation, high degree of automation, good reliability, suitable for long-term operation, etc. A true high-performance chromatograph.
1. Adopts a 7-inch color touch screen, the interface supports Chinese and English display,
2. All parameters of the instrument can be set and controlled through the touch screen and counter control workstation;
3. With RS232 communication port and LAM network port, adopts advanced 10/100M adaptive Ethernet communication interface and built-in IP protocol, which is convenient for establishing local area network and Internet to realize long-distance data transmission;
4. Internally design 3 independent connection threads, see the connection to the local processing, unit supervisor and superior administrative department, which is convenient for the unit supervisor and the superior administrative department to monitor the operation of the instrument and analyze the data in real time;
5. The special workstation equipped with the instrument can support the access of multiple chromatographs at the same time to realize data processing and reverse control of the instrument;
6. Modular design, easy to upgrade. Optional multiple detectors to meet complex sample analysis;
7. It realizes self-protection of gas circuit failure, automatic ignition, key point of flameout, automatic opening of gas circuit, and one-key start. Start the program during design, which can easily complete the on-line analysis of gas and liquid samples (equipped with sampling parts);
8. The system is designed with an automatic sampler interface, with multiple built-in drivers, and the automatic sampler can be loaded at any time;
9. The signal amplifying board has been specially treated and is always stable in harsh environments. The detector uses a logarithmic amplifier without flat head peaks, which ensures the accuracy of the content;
10. With complete audit trail, user authority management, electronic signature and other functions, the analysis data and documents are in line with GMP management requirements.
1. Oven Furnace size: 240×240×297mm Operating temperature range: above room temperature to 450°C
2. Injection port A variety of injectors can be equipped: packed column injector, split splitless injector, All inlets include electronic flow control modules with high precision electronic pressure/flow control; Use temperature 420℃; Column head pressure setting range: 1-100psi; Column head pressure control setting accuracy: 0.01psi; Total flow setting range: 0 - 1000ml/min (helium); 0 - 200ml/min (nitrogen); Flow setting accuracy: 0.1ml/min; Split ratio: 1:1000
3. Temperature control Temperature control range: 4°C~420°C above room temperature (1°C increment) Temperature control objects: oven, detector, sampler, auxiliary, a total of 6 channels The number of steps to ascend: 20 steps Rise rate: 0.1℃~50℃/min (increment 0.1℃) Constant temperature time: 0~9999.9min (increment 1min) Accuracy: ±0.01℃
4. Detectors parameters Hydrogen flame ionization detector (FID) ● Detection limit: Mt≤3×10-12g/s (n-hexadecane); ● Noise: ≤0.02mv ● Drift: ≤0.05mv/30min ● Linear range: ≥106
Thermal Conductivity Detector (TCD): ● Sensitivity: S≥8000mV•ml/mg (benzene-toluene) ● Baseline drift: ≤0.05mv/30min
● Noise: ≤0.02mv ● Linear range: ≥104
Electron Capture Detector (ECD): ● Detection limit: ≤1×10-14g/s ● Noise: ≤0.05mV ● Baseline drift: ≤0.2mV/30min ● Linear range: ≥103 ● Radioactive source: 63Ni
Flame Photometric Detector (FPD): ● Detection limit: (S)≤5×10-11g/s, (P)≤1×10-12g/s; ● Noise: ≤0.03mV ● Baseline drift: ≤0.2mV/30min ● Linear range: ≥103(S), 102(P)
Nitrogen Phosphorus Detector (NPD) ● Detection limit: ≤1×10-12g(n)/s (azobenzene) ● Noise: ≤0.03mV ● Baseline drift: ≤0.2mV/30min
|