HbA1c Analyzer Use High Quality Glycosylated Hemoglobin (HbA1c) Control LD-500 HbA1c Analyzer Use
Package Specification
There are two concentrations: Level 1 (low) and Level 2 (high).
Level 1 |
0.1mL×3 |
0.1mL×4 |
0.1mL×6 |
0.5mL×2 |
0.5mL×4 |
0.5mL×6 |
Level 2 |
0.1mL×3 |
0.1mL×4 |
0.1mL×6 |
0.5mL×2 |
0.5mL×4 |
0.5mL×6 |
Application
Control quality of glycosylated hemoglobin system in HbA1c analyzer.
Storage And Validation
- Stable at 2℃ – 8℃ for 12 months, direct exposure to sunshine avoided.
- Stable for 14 days at 2℃ – 8℃ after open-vial.
Attention: please stop using control if it is obviously contaminated.
Preparation For Use
- Open the vial carefully, avoiding any loss of material.
- Reconstitute by pipetting exactly 100ul of distilled water at +15°C to +25°C, into the vial.
- Close the rubber stopper and leave to stand for 10 to 15 minutes out of bright light before use.
- Swirl gently several times during the reconstitution period to ensure that the contents are completely dissolved.
- Prior to use, mix the contents by inverting the vial. Do not shake the vial as the formation of foam should be avoided. Ensure that no lyophilized material remains unreconstituted.
- Please close the rubber stopper after drawing the sample and put in the refrigerator soon to avoid containment and evaporation.
Cautions
- Store per request and use it within the validation. The reagent out of validation should be dealt with in experimental waste way.
- Different reagents with different lots can’t be cross used.
- Reagent box is for IVD use only and obey the laboratory SOP.