HbA1c Analyzer LD-100 Hemoglobin Testing Analyzer POCT One Step Detection Fast Testing Analyzer
Intended Use
- It is intended for the determination of percent hemoglobin A1c (HbA1c) in human whole blood.
- It is a semi-automatic instrument used to analyze glycated hemoglobin.
- It can be used to detect of HbA1c in medical units at all levels.And it's very protable to carry, you can testing HbA1c anytime.
- It has fully automatic system just need one step operation then the test will be start.
Specifications
Model | LD-100 |
Sample Type | Whole blood |
Sample Volume | 4ul |
Test Time | Less than 3 minutes |
Data storage | 7000 |
Method | Fluorescent Immunoturbidimetry |
Operating Type | 1-Step Operation, Fully automatic system |
Measurement Range | HbA1c 4.0~15%(NGSP Unit) eAG 68-385mg/dl(mmol/l) |
Operating Condition | Temperature:15~35℃ Humidity:15~80%R.H |
Storage Condition | Analyzer:10-60℃; Test kit: 2~8℃ |
Voltage | 12VDC-3.33A |
Analytical wave-length | 510nm and 580nm. |
Certification | NGSP, IFCC |

PRINCIPLE
- Uses the (borate affinity) fluorescent immunoturbidimetry
- Push a mixture containing a fluorescent borate conjugate into an integrated test cup and mix to dissolve sufficiently.
- The blood samples in the sample collector are then pushed into the buffer and mixed to release the hemoglobin from the red blood cells.
- The total concentration of hemoglobin depends on the initial reduced fluorescence signal.
- Fluorescent borate conjugates bind to glycosylated hemoglobin, and glycosylated hemoglobin is determined by detecting the decrease of fluorescent active ingredients.
- The percentage of glycosylated hemoglobin in total hemoglobin can be determined.
Feature
- Easy to operate, One Step Testing
- High Accuracy
- Cv<3%, Excellent Correlation With HPLC
- RFID For Batch Management
- Color touch screen
- External printer (Blue-tooth)
- USB Port, LIS Support
- One analyzer for testing without extra instrument
Reliable
IFCC calibrator | Normal | Abnormal | Normal | Abnormal |
Mean HbA1c% | 5.4 | 9.6 | 5.3 | 9.2 |
CV (%) | 2.5% | 1.9% | 2.4% | 1.8% |
Use Step
- Place cartridge in analyzer
- Push bead into reagents
- Insert sample into cartridge

Performance Parameters
- Accuracy: Is tested with the reference material as samples and the relative deviation of the determination results of the analyzer should be within the range of ±8%.
- Precision:Samples with a concentration of 4.0%-6.5% are detected, sample coefficient of variation (CV) of the test results is not more than 3%.
- Linearity:In the range of 4%-16% of HbA1c, the linear correlation coefficientrof the test results should be no less than 0.9900.
- Stability:Within 8 hours after the instrument is turned on and stable, the relative deviation of the test results is not more than ±3%.
Certificated