Medical device flow tester
Product introduction:
The medical device flow tester is designed according to the relevant provisions of the national standard for medical devices, which is used to detect the flow and flow rate of infusion set, blood transfusion set, infusion needle, anesthesia bag, catheter, etc.
The medical device flow tester is a special equipment designed according to the relevant provisions in the national standard for medical devices. Configuration requirements: main engine + 6m / M standard gas pipe 6m 2.3 pressure output range: higher than local atmospheric pressure, 10KPA ~ 200KPa, display mode: LED digital display; pressure output value can be set arbitrarily, the calibration error is not more than ± 2.5% of the reading. Test duration: 5 seconds ~ 99.9 minutes, LED digital display, the error is not more than ± 1s. It is suitable for infusion set, blood transfusion set, infusion needle, catheter, anesthesia filter, etc.
When the test is conducted according to Appendix B of YY 0285.5, the flow rate of pipes with nominal outer diameter less than 1.0 mm shall be 80% - 125% of the value marked by the manufacturer; for pipes with nominal outer diameter greater than or equal to 1.0 mm, the flow rate shall be 90% - 115% of the value marked by the manufacturer
11 kg net weight
Overall dimension: 400 × 290 × 240 (mm)