Internal diameter of the combustor | 100mm |
Height of the combustor | 450mm |
Pipe hole diameter of the long pole igniter trailing end diameter | 2±1mm |
Flowmeter | 160~1600L/Hadjustable |
Precision of the pressuregauge | level 2.5 |
Resolution of the pressure gauge | 0.01Mpa |
Resolution of the Oxygen concentrate indicator | 0.1% |
Oxygen sensor | imported from Germany with high precision |
Gas source | the GB3863 specified oxygen, the GB3864 specified nitrogen |
Experiment environment temperature, humidity | 10~35℃,45-75% |
Input pressure | 0.2-0.3Mpa |
Working pressure | 0.05-0.15Mpa |
Sample type | self-supporting materials and non-self- supporting materials |
Overall dimensions | 700mm(W)×400mm(D)×600mm(H) |
Power supply | AC220V, 50/60Hz 1A |
Air source source | oxygen, nitrogen( prepared by users) |
ASTM D 2863, ISO 4589-2, ISO 4589-3, NES 714, NES 715
Tips: Oxygen index tester requires not less than 98% of each industrial-grade oxygen/nitrogen bottle as the gas source. Because the above gas is a high-risk transport product, it cannot be provided as an oxygen index analyzer fitting, and it can only be used locally by users. Gas station purchase. (To ensure the purity of the gas, please purchase it at the local regular gas station)
What is oxygen Index test?
Measuring the Minimum Oxygen Concentration to Support Candle-Like Combustion of Plastics (Oxygen Index) ASTM D2863. Scope: ASTM D2863 is a method to determine the minimum concentration of oxygen in an oxygen/nitrogen mixture that will support a flaming burn in a plastic specimen.