Cone Calorimeter
Application
Suitable for testing the heat release rate of building materials.
Standard: ISO5660-1:2002
Technical Parameters
Power supply voltage | AC220V AC 50Hz |
Maximum operating power | ≥5KW |
Cone heater power | ≤5KW |
Heat output heat | 0~120KW/m2 |
The maximum placement of the sample box | 100mm × 100mm × 50mm |
Exhaust flow rate | 0.012m3/s~0.035 m3/s (adjustable) |
Timing value resolution | 1s error <1s/n |
Oxygen analysis | paramagnetic oxygen analyzer, concentration range 0-25% |
Smoke density analysis | (optional) |
Infrared analyzer | CO: 0-1% CO2: 0-10% (optional) |
Main feature
1.The test result data of SZL-1 cone calorimeter includes the following:
a Ignition time
b Average heat release rate within 180s after ignition for 180s
c Total heat release (mj/m2)
d Initial mass and residual mass (Kg) of the sample
2. f Average mass loss (kg/s) after ignition to the end of the test
g Radiation intensity (KW/M2) and displacement (m3/s)
h Heat release rate - time curve
i Smoke release rate (optional)
j Toxic gas (carbon monoxide, carbon dioxide) (optional) generation rate (optional)
High voltage spark generator: automatic positioning ignition
Modular combination of equipment: Users can select the standard configuration according to the actual application;
Equipment dimensions: length × width × height (1.68 × 0.686 × 1.92) m;
Equipment weight: 350 Kg.