The small content of ozone in the atmosphere is the main factor of rubber cracking. The ozone aging chamber simulates and strengthens the ozone conditions in the atmosphere, studies the action law of ozone on rubber, quickly identifies and evaluates the methods of rubber anti-ozone aging performance and anti-ozone agent protection efficiency, and takes effective anti-aging measures to improve the service life of rubber products.
This machine is used to test the cracking resistance of rubber products under the ozone environment. The ozone resistance tester can control and adjust the ozone environment of concentration to determine the ozone resistance of the test piece.
The ozone gas produced by the equipment from the silent high-efficiency ozone generator and a certain amount of treated air are mixed in the exchanger and then sent into the test chamber. The ultraviolet detector will feed back the ozone concentration continuously detected online to the ozone generator control system, so that the control system can effectively control the occurrence of ozone according to the set value, so that the test chamber can maintain the appropriate ozone concentration and meet all test conditions when the heating and humidification system works
JIS K 6259 ASTM1149 ASTM171 ISO1431 DIN53509 GB/T13642
GB 2951 (ozone resistance test method for wires and cables) and GB/T 7762-2003 (vulcanized rubber or thermoplastic rubber-ozone cracking resistance-dynamic tensile test).
Widely used in measurement and quality inspection; Rubber and plastic; Automobile production; Wire and cable; packaging; Instruments and Apparatuses; medical apparatus and instruments; Civil nuclear energy; Civil aviation; Colleges and universities; Research Institute; It is used for quality inspection of rubber printing products such as commodity inspection arbitration and technical supervision departments.
The length of the sample is basically parallel to the direction of the air flow when the sample is in contact with the ozonated air when the two ends of the sample are fixed on the rotating aluminum fixture under the specified elongation. Before the adjustable travel range of the sample is 0-40 mm, the sample rotates at a speed of (20~25) mm/s, so that each sample moves continuously along the same path in a plane perpendicular to the air flow. The time for the same sample to rotate for one cycle is (8~12) min, and the scanning area of the sample is about 40% of the effective area of the test chamber.
Model | SL-E20-02 [dynamic] |
Ozone concentration range | 10~500pphm/1000pphm/2000pphm adjustable |
Humidity range | the standard model does not contain humidity environment;if it needs to be increased;it needs to be discussed separately |
Temperature | RT+10~90 ℃ |
Humidity | 40% - 98% R.H |
Air velocity in the box | over 2 feet/sec |
Ozone concentration fluctuation | 10% |
Box capacity | 500 × 500 × 600mm |
Inlet air flow | A.0-6 cfm; B. 10-20 cfh |
Ozone concentration fluctuation | 10% |
Weight | 250kg |
Power supply | 1 ∮ |
AC220V | 15A |
Attachment | dynamic fixture 6 groups |