Mask Flame Retardant Performance Tester
APPLICATION:
Mask Flame Retardant Performance Tester is used for testing the anti-combustion performance of surgical masks
Standard:
GB19083-2010 technical requirements for medical protective masks
YY 0469-2011 medical surgical mask
Technical Parameters:
Movement speed of head mold | (60±5) mm/s |
Flame height | adjusted at 40±4mm |
Diameter of the flame temperature measuring probe | 1.5mm |
Temperature of the flame at (20±2) mm | (800±50) ℃ |
Distance between the top of the burner and the bottom of the mask | (20±2) mm |
Height adjustment range of burner | 30mm |
Head die | flame resistant and combustion resistant material |
Ignition time and smouldering time timer | 0 ~ 99.99 seconds, accuracy: ±0.01 seconds |
Machine dimension | 720mm×570mm×1000mm |
Power supply | Ac220V 50Hz 20W |
weigh | about 35Kg |
Features:
The ignition device is automatic, the flame capacity can be adjusted according to requirement