Specifications
Brand Name :
LEEB
Model Number :
Leeb250A
Certification :
CE, FCC EAC & ISO
Place of Origin :
China
MOQ :
1 Set
Price :
Based on quantity
Payment Terms :
T/T, PayPal, Western Union, L/C
Supply Ability :
500 sets per month
Delivery Time :
In 7-10 work days
Packaging Details :
Equipment Box
Name :
film thickness gauge
Application :
on Ferrous metal
Principle :
Magnetic Method
Shell :
Plastic
Probe :
Settled
Dimensions 115×55×23mm :
115×55×23mm
Description

Settled Fe Probe Coating Thickness Gauge Film Thickness Tester

Settled Fe Probe And Plastic Shell Coating Thickness Gauge With Switch off automatically or manually


Functions & Features

● High quality metal probes.
● Two measuring methods:continuous and single;
● Two working mode: direct and batch;
● Limit setting function.
● Switch off automatically or manually.

Measuring Materials
Magnetic Induction (Fe): Measuring the thickness of Non-magnetic coating on magnetic metal substrate,
such as aluminum, chromium, copper, zinc, rubber, paint on the base of steel, iron, alloy and magnetic
steel


Technical Parameters

Technical Parameters
Model No. Leeb250A
Measuring principle Fe
Measuring range (µm) 0~1250μm
Probe Settled
Shell Plastic
Accuracy ±(1~3%H+1) μm; H refers to the thickness of testing piece
Minimum resolution (µm) 0.1μm
Min curvature of the min area (mm) Convex1.5 Concave9
Diameter of the min area (mm) Φ7
Critical thickness of substrate (mm) 0.5
Memory 200 groups measured data
Dimensions 115*55*23mm
Power supply AAA Alkaline battery
Standard Configuration
Main unit,5 calibration specimens (48.5μm,99.8μm,249μm,513μm 1024μm),1 probe & substrate
Optional Accessories
Probes,Specimens,1 leather sheath

Troubleshooting
The following error information table to tell you how to identify and troubleshoot:
Error code
Implications of error code
The cause and solution
E02
Probe or instrument damage Repair probe or instrument
E03
Probe or instrument damage
Repair probe or instrument
E04
Measured values occur large fluctuations
(For example when soft layer measurements);
Magnetic field effect
When the measure menton
the softcover, auxiliary device to
measure should be adopted;Stay away from the
strong magnetic field environment
E05
Too close to the probe from the metal substrate when the phone is switched on
Probe from the metal substrate
E08
Probe or instrument damage
Repair probe or instrument
E11
Probe model in conformity with this group of the original data corresponding to the probe model
Replace the suitable probe
Choose an unused group units
Delete after calibration again
E15
Zero deviation is too big, can not be zero
Select the appropriate substrate or repair equipment
E20
This set of unit for calibration value
Choose an unused group unit Or delete after calibration again



High quality Equipment Box:

Settled Fe Probe Coating Thickness Gauge Film Thickness Tester

Send your message to this supplier
Send Now

Settled Fe Probe Coating Thickness Gauge Film Thickness Tester

Ask Latest Price
Brand Name :
LEEB
Model Number :
Leeb250A
Certification :
CE, FCC EAC & ISO
Place of Origin :
China
MOQ :
1 Set
Price :
Based on quantity
Contact Supplier
Settled Fe Probe Coating Thickness Gauge Film Thickness Tester

Chongqing Leeb Instrument Co.,Ltd

Active Member
5 Years
chongqing, chongqing
Since 2001
Business Type :
Manufacturer, Seller
Total Annual :
7500000-12000000
Employee Number :
100~200
Certification Level :
Active Member
Contact Supplier