Defibrillator | ||
Category | Specification | |
Output | Biphasic Truncated exponential | |
Energy Sequence | 150,150,200J | |
Charge Time | 8 sec.to 150J 10 sec.to 22J(New battery) | |
Analysis Time | 9 sec. | |
The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery | Less than 20 second | |
The maximum time from the initiation of rhythm analysis to readiness for discharge after 6 shocks | Less than 25 seconds | |
The maximum time from initially switching power on to readiness for discharge | Less than 30 seconds | |
Audible Prompts | 22 audible prompts | |
Visual Prompts | LED prompts: Automatic Self-test | |
Controls | Two buttons. On/Off, Shock | |
Output Energy Accuracy | ±15% into any impedance from 25 to 175Ω | |
The Maximum Voltage | 1200t50V | |
Output disabled when the patient impedance is outside limits | 20Ω to 200Ω |