Product Introduction:
UV-1800PC Double beam UV-Vis spectrophotometer, the successful implementation of the stringent requirements of high precision and high reliability of the measurement by computer on-line, available in the field of life sciences, agriculture, animal husbandry and fishery, geological exploration, food safety, environmental monitoring in
Specifications:
Net weight: 28Kg
Stray light: ≤0.03%τ(220nm NaI,340nm NaNo2)
Instrument Size: 560×450×230mm
Display: 6 inches high brightness LCD display
Stability: 0.0005/h(500nm Preheated)
Transmittance accuracy: ±0.3%τ (0-100%τ) ±0.002A(0~0.5A) ±0.003A(0.5A~1A)
Transmittance Repeatability: ±0.15%τ (0-100%τ) ±0.001A(0~0.5A) ±0.0015A(0.5A~1A)
Detector: Imports of silicon photodiode
Source: Imported deuterium lamp, tungsten lamp imports
Spectral Bandwidth: 1.8nm
Photometric range: -4~4A
Power: 120W
Power supply: AC 220V/50Hz,110V/60Hz
Metering: Automatic
Wavelength accuracy: ±0.3nm
Wavelength Reproducibility: ≤0.1nm
Wavelength tuning: Transmittance, absorbance, concentration, energy
Wavelength range: 190~1100nm
About Image:
● Automatic design, to achieve the most simple means of measurement |
● Rich measurement methods, with a wavelength scan, time scan, multi-wavelength determination, multi-order derivative determination (optional), dual-wavelength, three wavelength (optional) DNA protein measurement (option) and other measuring methods, can meet different measurements requirements, and can be displayed directly on a 6-inch screen |
● wide wavelength range, which can meet various areas of the wavelength range of requirements |
● can be powered down to save the measurement parameters and data, user-friendly |
● Measurement data can be output through the printer with USB interface |
● Realize spectral scanning by PC software control, more precise and flexible measurement requirements |
● based on user requirements optional hole rack, manual four with rack, manual Eighth rack, automatic Eighth frame, glass frame, test tube rack, 1cm colorimetric frame, 5cm colorimetric frame, 10cm colorimetric racks |
● 1.8nm spectral bandwidth can be customized according to user requirements to install to meet the stringent requirements of pharmacopoeia |
● Improved optimization of optical design, import source and receiver creating a system of high performance and high reliability |
● UV-1800PC series dual-beam optical system, successfully achieved the perfect combination of high precision and high reliability of the measurement can meet the requirements of various applications, can be used in biological research, biotechnology industry, pharmaceutical analysis, pharmaceutical, teaching and research, environmental protection , food hygiene, clinical testing, health and epidemic prevention and other areas |
● LSI design greatly improves the scalability and reliability of the system |
Applications:
Drug analysis: | National Formulary" Available UV-visible spectrophotometry drugs for drug testing, drug analysis, pharmaceutical and other industries |
Life Sciences: | DNA, protein concentration |
Animal husbandry and fishery: | Pesticide residues, crop composition testing, veterinary testing, feed testing, fertilizer testing, soil composition testing, testing and other aquaculture |
Metal elements: | Ag Cu Pb Zn Co Ni K Na Ca Mg Li Mn Cr As, including precious metals |
Geological Survey: | Determination of metal elements in minerals and inorganic salts |
Food safety: | Content of additives, preservatives, perfumes, fats, enzymes, sugars, spices, minerals, vitamins, etc. |
Environmental monitoring: | Monitoring water, air, rainfall and soil, etc., in which the determination of various pollutants |