UV-1700 has successfully achieved the strict requirements of high precision and high reliability measurement, which can meet the requirements of various applications, and can be used in biological research, biological industry, drug analysis, pharmaceutical, teaching and research, environmental protection, food hygiene, clinical testing, health and epidemic prevention and other fields. |
Wavelength Range: 190~1100nm Band Width:1.8nm , 1.0nm,or 0.5 , 1, 2, 4nm adjustable Wavelength Accuracy:±0.3nm Photometric Accuracy:±0.3% τ(0-100%τ) ±0.002A(0~0.5A) ±0.003A(0.5A~1A) Photometric Repeatability:0.15% τ(0-100%τ) ±0.001A(0~0.5A) ±0.0015A(0.5A~1A) Stray Light:≤0.03% τ (220nm NaI,340nm NaNO2) Stability:0.0005A/h@500nm Baseline Flatness: ± 0.001A Photometric Mode:T,A,C,E Wavelength Setting: Automatic Display Mode:6 inches high brightness blue LCD Light Source:Import Deuterium Lamp & Tungsten Halogen Lamp Power Requirement:AC 220V/50Hz 110V/60Hz |
1 | .Advanced voltage stabilizing power supply matching technology, the lamp source voltage of the machine is always kept in a stable state in the operation process, which extends the life of the lamp source and increases the accuracy of the test. |
2 | This product uses a double exhaust system, which can timely and efficiently discharge the heat generated in the machine operation, slow down the aging process of the machine, and continuously maintain the stable test state of the machine. |
3 | The machine adopts an integrated shock-proof fastening chassis, the chassis is installed with 20mm cast aluminum plate, the optical devices are fixed on the same platform, to ensure the consistency of the optical path, and effectively improve the stability of the instrument and the accuracy of the reading. |