Inductively Coupled Plasma Optical Emission Spectrometer ICP-6800
Inductively coupled plasma emission spectrometer(ICP6800) developed by Macylab Instruments with many years of technology accumulation. It is used to measure traces and traces of various substances (dissolvable in hydrochloric acid, nitric acid, hydrofluoric acid, etc.),and the content of metal elements or non-metallic elements. It is stable and reliable and with high degree of automation and easy operation.
At present, instruments are widely applied in various fields such as rare earth, geology, metallurgy, chemical industry, environmental protection, clinical medicine, petroleum products, semiconductor, food, biological samples, criminal science, agricultural research and so on.
Main features
Performance stability
The use of all-solid-state rf power supply, with small size, high efficiency, stable output power, with a variety of protection functions and many other advantages, the load using fully automatic matching technology, matching speed, improve the efficiency of the use of power supply and the stability of the instrument, and make the whole ignition process simple and convenient.
Injection automation
Adopting four-channel automatic design, the rotational speed can be set up according to the test demand to adjust the flow rate, carrier gas, plasma gas, auxiliary gas are all controlled by advanced mass flow controller, at the same time can be equipped with imported high salt nebulizer, imported hydrofluoric nebulizer and so on to meet the customer's various test needs.
Precision analysis
The middle step grating prism cross dispersion mode is adopted, there is no moving optical element, the ultra low stray light design is combined with the unique optical design, nitrogen distributed blowing, imported optical elements, intelligent and accurate automatic wavelength calibration algorithm.
Quick test
In American thermoelectric CID detector, the 160-1000nm range is continuously covered, and the intensity integral values of all analytical spectral lines are read out at one exposure to speed up the analysis.
Technical parameters
RFPower technical parameter
Circuit type | solid-state RF Power supply, with function of automatch |
Frequercy | 27.12MHz±0.05% |
Frequency Stablity | <0.1% |
Power Output | 800W-1500W |
Power Output Stability | <0.3% |
Escaped RF radiation | 30cm away from the instrument, electric fiekd: E<2Vtm |
Sampling System Technical Parameter
Output working coil imer diameter | 25mm |
Torpue tube | Three concentric, extemal diameter 20mm |
Coaxial nebulizer | Outer dameter 6mm |
Double barrel atomizing chamber | Outer diameter 34mm |
Gas Flow Controls
Plasma Argon Flowmeter | (100-1000)L/h(1.6-16L/min) |
Auxiliary Argon Fiowmeter | (100-1000)L/h(0.16-1.6L/min) |
Auxiliary Carrier Argon Flowmeter | (100-1000)L/h(0.16-1.6L/min) |
Pressune Maintaining Value | (0-0.4MPa) |
Cooling Water:Temperature | 20-25ºC,Rate of Flow>5L/min,Hydraulic Pressure>0.1MPa |
Spectrometer
Optics | Czerny-Turner type |
Focal length | 1000 mm |
Grating | Ion Beam Etching Holographic Grating, 3600L/mm or 2400L/mm |
Reciprocal linear dispersion | 0.26nm/mm |
Resolution | ≤0.007nm(3600 line grating); ≤0.015nm(2400 line grating) |
Wavelength range | 3600 line grating:(190nm~500) nm; 2400 line grating:(190nm~800) nm |
Minimum pace of stepping motor | ≤0.0006 nm |
Exit Slit | 12μm |
Entrance Slit | 10μm |
Photoelectric Converter Performance
Photomultiplier tube specification | R293/R928 |
Negative HV on PMT | 0-1000V |
Stability | <0.05% |
Overall Performance
Wavelength range of scanning: 195nm~500nm(3600L/mm line grating);195nm~800nm(2400L/mm line grating)
Repeatability: RSD(Relative Standard Deviation)≤1.5%
Stability: RSD(Relative Standard Deviation) ≤2%
Detection Limits (µg/L):
Element | WL(nm) | Limit | Element | WL(nm) | Limit | |
La | 408.672 | <3.0 | Cr | 267.716 | <5.0 | |
Ce | 413.765 | <5.0 | Al | 396.152 | <5.0 | |
Pr | 414.311 | <5.0 | Zr | 343.823 | <5.0 | |
Nd | 401.225 | <5.0 | Ag | 328.068 | <3.0 | |
Sm | 360.946 | <10.0 | Sr | 407.771 | <1.0 | |
Eu | 381.967 | <1.0 | Au | 242.795 | <5.0 | |
Gd | 342.247 | <10.0 | Pt | 265.945 | <5.0 | |
Tb | 350.917 | <3.0 | Pd | 340.458 | <5.0 | |
Dy | 353.170 | <3.0 | Ir | 224.268 | <10.0 | |
Ho | 345.600 | <3.0 | Rh | 343.489 | <10.0 | |
Er | 337.271 | <3.0 | Ru | 240.272 | <5.0 | |
Tm | 313.126 | <3.0 | Ba | 455.403 | <1.0 | |
Yb | 369.419 | <1.0 | As | 228.812 | ≤15 | |
Lu | 261.541 | <3.0 | Sb | 206.833 | ≤15 | |
Y | 371.030 | <1.0 | Bi | 223.061 | ≤10 | |
Sc | 335.373 | <1.0 | Hg | 253.652 | ≤15 | |
Ta | 226.230 | <5.0 | Pb | 220.353 | ≤15 | |
Nb | 313.340 | <5.0 | Ga | 294.364 | ≤10 | |
Mn | 257.610 | <3.0 | Se | 203.985 | ≤10 | |
Mg | 279.553 | <1.0 | Sn | 242.949 | ≤20 | |
B | 249.773 | <10.0 | Te | 214.281 | ≤10 | |
Zn | 213.856 | <3.0 | Ta | 226.230 | ≤5.0 | |
Co | 228.616 | <3.0 | Th | 283.730 | ≤10 | |
Si | 251.611 | <10.0 | Tl | 276.787 | ≤30 | |
Ni | 232.003 | <5.0 | Re | 227.525 | ≤5 | |
Cd | 226.502 | <3.0 | Ge | 209.426 | ≤15 | |
Fe | 239.562 | <3.0 | Os | 225.585 | ≤1 | |
Ca | 393.366 | <1.0 | W | 207.911 | ≤10 | |
Mo | 281.615 | <5.0 | Cu | 324.754 | <3.0 | |
V | 310.230 | <5.0 | Li | 670.784 | ≤3 | |
Be | 313.041 | <1.0 | Na | 588.995 | ≤20 | |
Ti | 334.941 | <3.0 | K | 766.490 | ≤60 |
Application Field
Certificate
Company profile
Macy(China) Instruments Inc is part of Shanghai Macylab Instrument Inc, is a high-tech enterprise with independent intellectual property rights . Macy hold the pioneering concept "technology - because you change " , and as a business purpose, and constantly explore, bold innovation. Especially in the field of analytical instruments ,Macy constantly develop advanced products that enable Macy to become a high-quality instrument resource supplier and leader.