Inductively Coupled Plasma Optical Emission Spectrometer ICP-6800
Main Features
Performance stability | The use of all-solid-state rf power supply, with small size, high efficiency, stable output power, with a variety of protection functions and many other advantages, the load using fully automatic matching technology, matching speed, improve the efficiency of the use of power supply and the stability of the instrument, and make the whole ignition process simple and convenient.
|
Injection automation | The instrument is equipped with a four-channel twelve-roller high-precision peristaltic pump, which can guarantee the precision of injection and prevent liquid accumulation at the same time. The speed of the peristaltic pump can be adjusted continuously to meet various test requirements of customers. |
Flow control parameters | Adopt imported high precision mass flow controller. |
High precision constant temperature system | The whole optical path of adopt the precision constant temperature protection system. The temperature control can be set in real time according to the actual ambient temperature of the customer's laboratory, without the need to turn on the air conditioner continuously for a long time. Temperature control precision ≤±0.1%, precision constant temperature protection system to ensure the stability of the optical path and makes the test data more stable. |
RFPower technical parameter:
Circuit type :solid-state RF Power supply, with function of automatch
Frequercy :27.12MHz±0.05%
Frequency Stablity :<0.1%
Power Output :800W-1200W
Power Output Stability :<0.3%
Escaped RF radiation :30cm away from the instrument,
electric fiekd: E<2V/m
Sampling System Technical Parameter
Output working coil imer diameter :25mm
Torpue tube :Three concentric, extemal diameter: 20mm
Coaxial nebulizer :Outer dameter 6mm
Double barrel atomizing chamber :Outer diameter 34mm
Gas Flow Controls
Plasma Argon Flowmeter :(100-1000)L/h(1.6-16L/min)
Auxiliary Argon Fiowmeter: (10-100)L/h(0.16-1.66L/min)
Auxiliary Carrier Argon Flowmeter :(10-100)L/h(0.16-1.66L/min)
Pressune Maintaining Value :(0-0.4MPa)
Cooling Water:Temperature :20-25ºC,Rate of Flow>5L/min,Hydraulic Pressure>0.1MPa
Spectrometer
Optics :Czerny-Turner type
Focal length :1000 mm
Grating :Ion Beam Etching Holographic Grating, 3600L/mm or 2400L/mm
Reciprocal linear dispersion :0.26nm/mm
Resolution :≤0.007nm(3600 line grating); ≤0.015nm(2400 line grating)
Wavelength range :;3600 line grating:(190nm~500) nm; 2400 line grating:(190nm~800) nm
Minimum pace of stepping motor :≤0.0006 nm
Exit Slit :12μm
Entrance Slit :10μm
Overall Performance
Wavelength range of scanning | 195nm~500nm(3600L/mm line grating);195nm~800nm(2400L/mm line grating) |
Repeatability | RSD(Relative Standard Deviation)≤1.5% |
Stability | RSD(Relative Standard Deviation) ≤2% |
Detection Limits (µg/L): |