S6100A Anesthesia System For Adult and Pediatric
Description & Application
The S6100A anesthesia machine is a highly reliable and versatile device suitable for various healthcare settings, including Intensive Care Units (ICU), operating rooms, and Anesthesiology Departments. It is specifically designed to ensure ease of use and maximum patient safety during anesthesia procedures. The machine incorporates essential functions and features advanced ventilation modes for effective respiratory management in adults, children, and infants. With a focus on professional design and incorporating proven ventilation technology, the S6100A anesthesia machine offers seamless integration with an advanced anesthesia table, optimizing workflow efficiency for healthcare professionals. Its ergonomic design and user-friendly interface make it an indispensable tool in anesthesia practice, delivering optimal patient care and outcomes.
Advantages
The S6100A anesthesia machine offers advanced ventilation modes, prioritizes patient safety, and has a user-friendly design. It is versatile for patients of different age groups and integrates well with other medical devices. The machine is durable and reliable, providing optimal care during surgical procedures.
Trust point
- Providing customers with high quality and cost-effective anesthesia machine.
- The Medical team is a group of experienced and dedicated professionals with a passion for more than 20 years. We listen to our customers and integrate their insights in our machine development. Flexible configurations to suit customers’ needs.
- Countless feedback from customers give good reviews.
- Real time pressure-time, flow-time loop Oscillogram and high precision O2 concentration detection function included.
Features
- 7" TFT LCD screen displays the Ventilation parameters, Alarm information and Oscillogram.
- High precision flowmeter, instantly know the fresh gas flow to your patient.
- Integrated breathing circuit design, ensure easy operating and keep tidy.
- Multiple working modes such as volume control and pressure limit, adapt to wide range patient.
- Vaporizer with temperature, pressure, flow compensation and self-lock function, keep safety anytime.
- Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects.
- Pressure-time, flow-time oscillogram show in real time.
- Vital sign monitor and Anesthetic gas monitor are optional.
- ACGO and fast oxygen supply - Emergency use and revival after opertation.
Ventilation mode |
IPPV, A/C, SIMV, SIGH, MANUAL |
Ventilator parameter range |
Flowmeter | O2 (0 ~ 10 L/min) N2O (0 ~ 10 L/min) or AIR (0 ~ 10 L/min) |
Rapid oxygen supply | 25 L/min - 75 L/min |
Tidal volume | 0, 20 mL ~ 1500 mL |
Maximum MV | ≥18 L/min |
Frequence(Freq) | 1 /min - 100 /min |
I:E | 2:1(1:0.5)- 1:6 |
PEEP | 0 cmH2O - 30 cmH2O |
Pressure triggering sensitivity (PTr) | -20 cmH2O - 0 cmH2O (Based on PEEP) |
Flow trigger sensitivity (FTr) | 0.5 L/min - 30 L/min |
Pressure control (PC) | 5 cmH2O ~ 70 cmH2O |
SIGH | 0 (off) 1/100 - 5/100 |
Apnea ventilation | OFF, 5s - 60s |
Pressure limit | 20 cmH2O - 100 cmH2O |
Alarm and protection |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | < 11.3 ± 0.3 V |
No tidal volume | ≤ 5 mL within 6 s |
High oxygen concentration alarm | 19 % - 100 % |
Low oxygen concentration alarm | 18 % - 99 % |
High airway pressure alarm | 2 cmH2O - 100 cmH2O |
Low airway pressure alarm | 0 cmH2O - 20 cmH2O |
High minute volume alarm | Adult (5 L/min ~ 20 L/min) Paed (1 L/min ~ 15 L/min, 0 ~ 10 L/min) |
Low minute volume alarm | 0 ~ 10 L/min |
Continuous pressure alarm | (PEEP+1.5 kPa) over 16 s |
Suffocation warning | 5 s - 60 s no spontaneous ventilation |
The maximum limited pressure | ≤ 12.5 kPa |
Fan error | Show on screen |
Oxygen deficit | Show on screen |
Monitoring parameter |
Frequence(Freq) | 0 /min - 100 /min |
Tidal volume (Vt) | 0 mL - 2000 mL |
MV | 0 L/min - 100 L/min |
Oxygen concentration | 15 % - 100 % |
Working condition |
Gas source | O2, N2O, AIR |
Pressure | 280 kPa ~ 600 kPa |
Voltage | 100 ~ 240 V |
Power frequency | 50/60 Hz |
Oscillogram |
P-T (pressure - time) |
F-T (flow - time) |
P-V loop (pressure - volume loop) |
Wooden case packing size |
Wooden case packing size : L 790 * W 740 * H 1460 mm |
G.W. :103 KG N.W. :76 KG |