Application Conditions and Requirement | |||||
Environment | Power Requirement | Safety | |||
Operation Temperature | Humidity | Altitude | Voltage Supply | IP Class | Conformity |
-10°C--+40°C | 35%--85% | < 2500m | 3x220/380v±10%(or customized) 50/60Hz±2Hz | IP 30 | CE conformity |
Type | YC1891D-6 | YC1891D-12 | YC1891D-24 | YC1891D-48 |
Number of meter positions | 6 | 12 | 24 | 48 |
Accuracy | 0.05 | |||
Power Source | YC8-101-250 | YCS-101-600 | YCS-101-1200 | YCS-101-2500 |
Standard Meter | SZ-01A-K3 | |||
MSTV | HJ200G-6 | HJ200G-12 | HJ200G-24 | HJ200G- 48 |
Scanning Head | YCG-2-series | |||
Voltage | ||||
Test voltage output (Phase-Neutral) | 24V-300V(or customized) | |||
Power of the voltage output | Max.100VA | Max.250VA | Max.500VA | Max.1000VA |
Resolution | Better than 0.01% of full scale value of range | |||
Setting accuracy | Better than 0.05% at the final range value | |||
Stability | Better than 0.005%/h(integration time 150 s) | |||
Load regulation from O-max.load | Better than 0.01% | |||
Distortion factor | Better than 0.3% for linear resistance load | |||
Harmonics | 2nd - 41st free programmable | |||
Current | ||||
Test current output | 1mA-120A/200A | |||
Power of the current output | Max.250VA Max.500VA Max.1000VA Max.2000VA | |||
Resolution | Better than 0.01% of full scale value of range | |||
Setting accuracy | Better than 0.05% at the final range value | |||
Stability | Better than 0.005%/h(integration time 150 s) | |||
Load regulation from O-max.load | Better than 0.01% | |||
Distortion factor | Better than 0.3% for linear resistance load | |||
Harmonics | 2nd - 41st free programmable | |||
Phase angle | ||||
Phase angle | 0°-360° | |||
Resolution | 0.01° | |||
Setting accuracy | 0.1° | |||
Frequency | ||||
Range | 45Hz-65Hz | |||
Resolution | 0.01Hz | |||
Error display | ||||
Type of error display | Red LED | |||
Resolution of the error displays | 8 digits |