YC-1893GT Rail-Mount Stationary Energy Meter Verification System
Device Features
- Flexible System Characteristics
- RJ-45 / RS-485 / CAN Interface
- Carrier / Remote Signal / State Grid Bluetooth Communication Interface
- Microcomputer Automatic Calibration Operation and Keyboard Manual Calibration Operation
- Flexible Calibration Plan Settings
- Various Report Formats that Can Be Modified According to User Requirements
- Communication Protocol for Smart Metering
- Barcode Input and Networking Functions
- Supports 645, 698 Communication Protocols
Technical Specifications
Device Accuracy Level: 0.05 Level / 0.1 Level
Output Voltage
- Range: 3x57.7V / 3x100V / 3x220V / 3x380V
- Adjustment Range: 0~120% Un
- Adjustment Precision: Minimum 1mV
- Waveform Distortion: ≤0.3%
- Phase Shift Impact: <0.1%
- Load Regulation: ≤±0.3% of Full Scale Value (Pure Resistive, Inductive, Capacitive)
- Monitoring Accuracy: Multi-functional Standard Meter Monitoring
Output Current
- Range: 3x0.02A~100A Automatic Range Selection (Power Source Part, Default Imax=50A)
- 3x0.2A~500A (Through 1:10 ICT Transformation Output. ICT Accuracy Level: 0.01.)
- CT Transformation Output, Starting Current: 200mA (Minimum), Starting Current Accuracy ≤3%
- Simulated Rogowski Coil Output Voltage: 0.16mV-400mV (Corresponding to Rail Meter 0.2A~500A Input)
- Adjustment Range: 0~120% In
- Adjustment Precision: Minimum 0.1mA (Power Source Part)
- Waveform Distortion: ≤0.3%
- Phase Shift Impact: <0.1%
- Load Regulation: ≤±0.3% of Full Scale Value (Pure Resistive, Inductive, Capacitive)
- Monitoring Accuracy: Multi-functional Standard Meter Monitoring
Output Phase
- Adjustment Range: 0.01°~360°
- Adjustment Precision: Minimum 0.01°
- Load Regulation: <±0.3°
- Monitoring Accuracy: Multi-functional Standard Meter Monitoring
Output Frequency
- Adjustment Range: 45.000Hz~65.000Hz
- Adjustment Precision: Minimum 0.001Hz
- Load Regulation: <±0.1%
- Monitoring Accuracy: Multi-functional Standard Meter Monitoring
Output Power Symmetry
- When the device indicates (or defaults to) symmetry, the actual output symmetry should not exceed the specifications in Table 3-3.
Output Power Stability
- After the device reaches a thermally stable state, within the specified output load range, the output power stability Py should not exceed the specifications in Table 3-4.
(Table 3-3)
Device Accuracy Level | Voltage Symmetry (%) | Current Symmetry (%) | Phase Symmetry (°) |
0.05 Level | ±0.5 | ±1.0 | 2 |
0.1 Level | ±0.5 | ±1.0 | 2 |
(Table 3-4)
Device Accuracy Level | 0.05 Level | 0.1 Level |
γР at 1.0 | 0.02 | 0.05 |
γР at 0.5L | 0.05 | 0.1 |
Timing Signal
- Modular Timing Frequency Meter, Output Timing Frequency 625kHz
- Internal Crystal Oscillator Frequency Accuracy ≤0.4x10^-6, Annual Stability ≤0.1x10^-6
- (The above are default configurations, for other performance parameters and functions, please contact our company for customization)