HS-3163P Portable single-phase test system is a high-performance calibrating instrument which is designed with international new digital technology. This test system consists of an integrated single-phase wide range voltage & current source and a single-phase electronic reference standard meter of accuracy class 0.05%. It is wide measuring range, 0.05% class high accuracy ,high tolerance to unwanted external influences and more utility functions.
HS-3163P is mainly used on-site to test energy meter’s installation,running and analyze power system.
Main technical features
Main measurement functions
Application Range
2,General Technical Index
Auxiliary power supply | 198 VAC … 253 VAC / 47 … 63 Hz |
Power consumption | 200 VA |
Shell | Hard plastic |
Dimension | 355 x 272 x 166 mm(W*H*D) |
Weight | ≤7KGS( not including spare parts) |
Working Temp. | -10 °C ~ +50 °C |
Storage Temp. | -20 °C … +60 °C |
R.H. | ≤85%,≤21°C |
3,Voltage source technical standard
Output voltage range/power | 20 V … 300 V / 20 VA | ||
Resolution | 0.001 V | ||
THD | ≤ 0.5% | ||
Stability | ≤ 0.05% (3 min.) | ||
Adjust range | 0% … 120%(240V) | ||
Frequency | range | accuracy | resolution |
45 Hz – 65 Hz | ±0.01 Hz | 0.01 Hz |
4,Current source technical standard
Output current range/power | 5 mA … 60 A / 50 VA | ||||
Current output(Terminal) | current | range | Smax / Umax | ||
0.1 A | 5 mA … 100 mA | 0.6 VA / 6 V | |||
1 A | 0.1 A … 1 A | 5 VA / 5 V | |||
10 A | 1 A … 10 A | 12 VA / 1.2 V | |||
50 A | 10 A … 50 A | 50 VA / 1 V | |||
Resolution | 0.01%*range | ||||
THD | ≤ 1% | ||||
Stability | ≤ 0.05% (3 min.) | ||||
Adjust range | 0% … 120% | ||||
Phase angle | range | accuracy | resolution | ||
0°…359.99° | 0.2° | 0.1° | |||
Frequency
| range | accuracy | resolution | ||
45 Hz – 65 Hz | ±0.01 Hz | 0.01 Hz | |||
5,Reference standard meter technical standard
5.1,Measurement range
Measurement | range | input/ sensor |
Voltage | 10 V … 300 V | U |
Current | 5 mA … 60 A | I |
10 mA … 120 A | clamp 20A,100A |
5.2,Voltage and current measurement accuracy
voltage / current | ≤±E[%]1 2 | |
Measurement | range | class |
Voltage | 100 V … 300 V
10 V … 100 V | 0.05
0.2 |
Current | 20 mA … 60 A
5 mA … 20 mA | 0.05
0.2 |
Clamp | 10% ≤ RG ≤ 120%
1% ≤ RG < 10% | 0.2
0.5 |
5.3,Energy measurement accuracy
Power/energy voltage:100 V … 300 V | ≤±E[%]1 2 3 | |
Measurement /output current | range | class |
Active power(P),apparent power(S) | ||
Output | 20 mA … 60 A
5 mA … 20 mA | 0.05
0.2 |
Clamp | 10% ≤ RG ≤ 120%
1% ≤ RG < 10% | 0.2
1.0 |
Reactive power(Q) | ||
Output | 20 mA … 60 A
5 mA … 20 mA | 0.1
0.2 |
Clamp | 10% ≤ RG ≤ 120%
1% ≤ RG < 10% | 0.4
1.0 |
Magnetic influence:≤0.07 % / 0.5 mT³(45 Hz … 66 Hz);
Temp. coefficient(TC): | |
range | ≤ ±TC[%/℃] 3 |
0 ℃ … +40 ℃ | 0.02 |
-10 ℃ … +50 ℃ | 0.05 |
5.4,Frequency,phase,PF measurement accuracy
F/Phase angle/PF | ≤ ±E | |
Measurement | range | |
F | 45 HZ … 65 Hz | 0.01 Hz |
θ | 0.00° … 359.99° | 0.1° |
PF | -1.000 … +1.000 | 0.002 |
remarks:
x.x: Related to measurement range(Max range,FS),E(M)=FS/M*x.x( for example: 0.2, FS =46 v, E(10V) = 46/10 * 0.2 = 0.92 %)
5.5,Pulse input/output
Input TTL | 3 … 6 VDC | |||||
Input F | Max.200 kHz | |||||
Output TTL | 5 VDC | |||||
Pulse width | ≥ 10μs | |||||
Constants:
Active power,reactive,apparent power [imp/kWh(kvarh,kVAh)] | C = 1×109/In’;In’= In×(100%,50%,25%)
Constants depends on internal ranges(In) Details see constants table | |||||
Current In[A] | ||||||
Measurement | 0.1 | 1 | 10 | 50 | ||
100%In | 50%In | 25%In | ||||
Output F | C’= C/3600000[imp/Ws(vars,Vas)]
fo = C’×P∑(Q∑,S∑) fmax = 16.666kHz | |||||
6,Others
Position | 1 |
COM. port | RS232,Bluetooth |
Energy error indication range | -99.999%~+2000.0% |
Other functions | Victor diagram display,error display,harmonic analysis display |