HS-6103F SINGLE PHASE ENERGY METER TEST BENCH
AUTOMATIC CHANGE TO CALIBRATE/TEST LINE AND NEUTRAL
(48 POSITIONS)
This specification for single phase 2 circuit loop energy meter test bench.
Voltage Source:
1. Output Voltage Range: 0- 275/300V
2. Output Max Power: 800VA
3. Resolution: 0.01%
4. Stability: Better than 0.05% / 3 Min(Time base 5S)
5. Distortion factor: < 0.5% at linear load;
6. Frequency range: 45Hz~65Hz;
7. Power efficiency: more than 85%;
8. Ambient temperature:10℃~50℃
9. Electronic overload and short-circuit protection.
Current Source:
1. Output current: range: 1mA~100A;+/-20%
2. Output maximum power: 2000VA
3. Resolution: 0.01%;
4. Stability:better than 0.05%/3 min(time base 5S);
5. Distortion factor: < 0.5% at linear load;
6. Frequency range: 45Hz~65Hz;
7. Power efficiency: greater than 85%;
8. Ambient temperature:10℃~50℃
9. Electronic overload protection and security for open current circuit.
Reference Meter:
HS-5100 Single Phase Reference Meter Accuracy class :0.05%
Accessories:
1. Phase Angle: 0°~360° , resolution of phase angle setting: 0.1°;
2. Output frequency range: 45Hz~65Hz; resolution: 0.01Hz
3. Output stability of power: <0.05% (180S)
4.. Input power: 220/240V +/-15%, 50/60Hz;
5. RS485 Communication Capacity.Universal Serial Bus
6. Current Connect Type: 1-2,4-3 (Dual Loop)
Dimensions:
Type of bench | Meters | Number | Number of row | Size |
Integration | 6 12 24 48 | 1 1 1 2 | Single Double Double Double | 1600×750×1100 mm 1600×750×1920 mm 2200×750×1920 mm 2000×750×1920 mm |
Fission | 12 16 24 48 | 1 1 1 2 | Single Double Double Double | 1600×700×1920 mm 2200×700×1920 mm 2200×700×1920 mm 2000×700×1920 mm |
Cabinet: 600×800×1920 mm |
QUICK CONNECTOR & SCANNING HEAD
FOR 2 CIRCUIT CURRENT LOOP TESTING(WITH PAIR CT)