HS-6103 Single-phase energy meter test bench. This Bench Comprises of PWM power source,standard digital energy meter,error processor,and PC processing software.Can test 3-60 Variors Energy Meter. Technical Parameters: 1.Accuracy Class:0.05/0.1/0.2 % 2.Standard Energy meter: HS-5100 Single-Phase Wide-measurement-range Energy Meter,Accuracy class:0.05/0.1/0.2 ; Multi-path Voltage transformer: Accuracy class:0.01. Rated load per path: 10 VA 3.Voltage Range(V): 220V(240V Option) Output Range: 0~125%,Resolution:0.01% full range 4.Current Range(A):0.1 ~100A(200A option) Output Range:0~120%,Resolution:0.0001A(Min) 5. Phase Angle:0.360.Resolution of phase angle setting:0.01 6. Frequency: 45Hz ~65Hz,Resolution:0.01Hz 7.Output waveform distortion:Both Voltage and Current 0.5% 8.Output Voltage,current and power stability: 0.05% /3min 9.Output Capacity: Current: 1000VA(120A) Voltage: 400 VA(24 meter testing stations) If the number of energy meter to be tested is increased whenever necessary,its maximum load may be extended to current:1800VA, and Voltage:1000VA 10.Load Characteristics: Resistive,inductive and Capacitive(less than 4uF) 11.Starting current output:5mA(Min);Accuracy:5%;Starting power:Accuracy 5% 12.Test Stations: 24/48 (Standard), also 3-60 option. 13.Testing function: Accuracy testing,Creep testing,Starting up testing,Harmonic testing,repeatability testing,influencing testing,dial testing etc. 14.Input power supply:220V AC 10%,50Hz;Max power consumption:1500VA(24 energy meter testing stations) 15. RS485/Rs-232 16. GPS timing system, HS-1012 standard clock,Option |