OPC-L2 Oil Particle Counter
OPC-L2 Portable Oil Particle Counter adopts Laser Extinction Technology, is designed to measure the particles’ size and quantity in oil liquid. It is Suitable for tests of pollution degree of solid particles in hydraulic oil, lubricating oil, fire resistant oil, insulating oil, turbine oil, organic liquid and polymer solution, etc. So our instrument can be used in aviation, spaceflight, electric power, petroleum, chemical industry, transportation, port, metallurgy, machinery, automobile manufacturing and many other fields.4.1 Power Choice: External power or Battery.
* Up to 64 testing channels can be set up at the same time according to customers' needs.
* Built-in data analysis system, a test can give all the standards of the test data and contamination levels
* Built-in viscosity, moisture and temperature sensor module, in the accurate testing of particle distribution at the same time to provide viscosity, moisture content saturation and ppm values and temperature reference values (optional)
* Any particle size can be set, with 9999 built-in particle sizes for easy particle size analysis.
* A variety of sampling containers such as standard sampling bottles or sampling cups can be used to meet the testing requirements of different industries
* Full-featured automatic color touch screen operation, support for Chinese and English dual interface, easy to operate
* Single-channel and multi-channel calibration can be carried out to realize the automatic calibration function.
* With RS232 interface, can be connected to a computer or laboratory platform for data processing.
* With massive data storage, printing function, can store 1000 groups of data and support U disk storage data.
OPC-L2 oil particle counter can be widely used in aviation, aerospace, electric power, petroleum, chemical industry, transportation, port, metallurgy, machinery, automobile manufacturing and other fields.
Light source | Semiconductor laser |
Particle size range | 0.8μm~600μm (calibrated according to 1~100um or 4~70um(c)) |
Detection channels | 8~64 channels optional, arbitrary setting of particle size size |
Sensitivity | 0.8μm (ISO4402) or 3μm(c) (GB/T18854, ISO11171) |
Resolution | <10% (GB/T18854, ISO11171) |
Repeatability | RSD<2% |
Sample detection viscosity | ≤650cSt (viscosity is too large can be combined with heating or dilution method detection) |
Sampling volume | 0.2~1000mL, interval 0.1mL |
Sampling accuracy | better than ± 0.5% |
Sampling speed | 5 ~80mL/min |
Pneumatic chamber | Positive and negative air pressure chamber device, to realize sample degassing and high viscosity sample detection |
Maximum vacuum of air pressure chamber | -0.08Mpa |
Maximum positive pressure of air pressure chamber | 0.8 Mpa |
Maximum particle concentration | 12000~40000 particles/mL |
Power supply | 110~245V AC, 50Hz, 70W |
Overall Dimension | 340mm*410mm*650mm |
Net weight | 25kg |