UVOZ-1100D ozone gas analyzer is designed and manufactured according to the verification and calibration method of the People's Republic of China-ultraviolet photometric method, and the Ministry of Construction of the Ministry of Construction of the People's Republic of China. The current ozone concentration is calculated by measuring the change of optical signal intensity before and after the absorption of ultraviolet light. Equipment using advanced double light road ultraviolet light source system, equipped with lamp intelligent management system, after boot can make the ultraviolet lamp quickly to measurement status, using foreign separation light pool technology, equipment has no leakage, high pressure resistance, large flow sampling gas impact, easy to clean, easy maintenance, simple operation, low use cost, etc. It is mainly used for continuous detection of ozone gas in various industrial environments, and can also be used for detection of ozone gas in pipelines, containers and other environments in operation. Widely used in pharmaceutical, chemical, municipal, sewage treatment and other industries, the ozone generator export concentration and ozone generator output measurement.
Test method: dual-light path ultraviolet light absorption method, long-life light source system, high measurement accuracy. Measurement principle: according to the Lambert-Beer law, accurate measurement through the luminosity absorption principle. Light source system: foreign long-life ultraviolet light source system, free warranty for 2 years. Usage mode: indoor installation and use, equipped with power supply and signal terminals. Optical pool system: separated optical pool technology, no leakage, high pressure resistance, large flow rate sampling gas impact. Intelligent compensation: built-in temperature and pressure compensation and display, with light source automatic compensation function. Operation mode: the user can select the manual zero calibration and automatic zero calibration according to the running status, and the automatic zero calibration time can be set. Display unit: ppm, mg / Nm 3 optional. Data display: hd color touch screen. Output function: 4-20 mA, RS485 communication, output of two groups of alarm points, etc. Standard configuration: ozone exhaust destroyer, air intake filter.
Model | UVOZ-1100D |
Measurement range | 0-100ppm; |
Optional range | 0-1000ppm; |
Light source life | about 20000-30000 hours; |
Sampling method | negative pressure sampling; |
Display interface | 4.3-inch touch screen operation interface; |
Display unit | ppm, mg/Nm3 optional; |
Display resolution | 0.01ppm; |
Sampling gas flow | 200-800mL/min (flow adjustable); |
Input pressure | atmospheric pressure 101KPa; |
Concentration error | Maximum 0.3%FS (intermittent zero sampling); |
Linearity error | Maximum 0.3%FS (intermittent calibration zero sampling); |
Zero drift | <±0.3%.FS; |
Response time | Signal 0.03 seconds, display data once every 5 seconds (intermittent calibration zero sampling); |
Ambient temperature | -20~50ºC; |
Piping interface | Φ6 quick-screw connector, (equipped with air inlet filter);. |
Bypass sampling calibre | Φ6 (6mm*4mm); |
Communication mode | RS-485; |
Output mode | 4-20mA (active); |
Relay signal | high alarm point relay signal, low alarm point relay signal; |
Power supply | 24V DC; |
Dimensions | 160mm (W) x 90mm (H) x 330mm (D); |
Ozone generator manufacturers,
Municipal water industry,
Industrial wastewater industry,
Fine chemical industry,
Food and drinking water industry,
Space disinfection industry,
Swimming pool disinfection industry,
Flavor synthesis industry ,
other industries using ozone generators.