Medical Refrigerating Freezer PCR Laboratory
1.Type and volume: vertical refrigerating freezer, the upper is for refrigeration and the lower for freezing. Refrigeration volume ≥185L, freezing volume ≥95L, and the refrigerator can be deactivated separately.
2.Structure: Both the refrigerator and the freezer require a hidden evaporator design without exposing the evaporator.
3.Temperature: 2~8℃temperature range in the refrigerator box, -20℃~-30℃adjustable in the freezer box
4.Shelf: A simple shelf structure is required in both the refrigerator and the freezer.
5.Energy conservation and environmental protection: The design of the refrigeration system should meet the requirements of environmental protection and energy conservation, and hydrocarbon refrigerant is required.
6.Alarm type: Over temperature alarm, sensor failure alarm, power failure alarm, low battery alarm, door opening alarm
7.Alarm mode: audible buzzer and flashing light alarm, with remote alarm interface
8.Refrigerating mode: air-cooled and frost-free design in the refrigerator, direct refrigerating design in the freezer
9.Condensation method: forced air circulation condensing
10.Lighting: There is LED lamp in the refrigerator box.
11.Safety: Both temperature zones are required to be equipped with independent locks, with at least 2 padlocks per lock.
12.Test hole: Both the refrigerator and the freezer are equipped with a test hole, the diameter of which is not less than 20mm, convenient for the use of cold chain monitoring probe line.
13.The product has a medical device registration certificate.